![Hanna Instruments HI 903 Quick Start Manual Download Page 149](http://html.mh-extra.com/html/hanna-instruments/hi-903/hi-903_quick-start-manual_590963149.webp)
A1-3
APPENDIX 1
A1
TECHNICAL SPECIFICATIONS
Range
100 ppm to 100%
Resolution
1 ppm to (0.0001%)
Result Units
%, ppm, mg/g,
g/g, mg,
g, mg/mL,
g/mL,
mg/pc,
g/pc
Sample Type
Liquid or Solid
Determination
Pre-Titration Conditioning
Automatic
Background Drift Correction
Automatic or User Selectable Value
Endpoint Criteria
Fixed mV persistence, Relative drift stop or Absolute drift
stop
Dosing
Dynamic with optional pre-dispensing
Result Statistics
Mean, Standard Deviation
Clip-Lock
TM
Exchangeable Burette System
Dosing Pump
Resolution
1/40000 of the burette volume (0.125
L per dose)
Accuracy
± 0.1% of full burette volume
Syringe
5-mL precision ground with PTFE plunger
Valve
Motor-driven 3-way, PTFE liquid contact material
Tubing
PTFE with light block and thermal jacketing
Dispensing Tip
Glass, fixed position anti-diffusion
Titration Vessel
Conical with operating volume between 50-150 mL
Solvent Handing System
Sealed system, integrated diaphragm air pump.
Electrode
Type
Dual platinum pin, polarization electrode
Connection
BNC
Polarization Current
1, 2, 5, 10, 15, 20, 30 or 40
A
Voltage Range
2 mV to 1000 mV
Voltage Resolution
0.1 mV
Accuracy
± 0.1%
Stirrer
Type
Magnetic, Optically regulated, digital stirrer
Speed
200- 2000 RPM
Resolution
100 RPM
External Stirrer
4-pin mini DIN Connection allows for the control of an
external stirring apparatus
Peripheral Devices
PC
Transfer methods and reports via USB connection to
a PC using the HI 900 PC Software
Summary of Contents for HI 903
Page 1: ...1 QUICK START GUIDE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 0 ...
Page 4: ...4 QUICK START GUIDE ...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View ...
Page 18: ...18 QUICK START GUIDE QS 903 02 11 ...
Page 19: ...1 INSTRUCTION MANUAL HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Page 22: ......
Page 24: ...1 2 INTRODUCTION ...
Page 26: ...2 2 SETUP ...
Page 30: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View ...
Page 40: ...2 16 SETUP ...
Page 42: ...3 2 USER INTERFACE ...
Page 50: ......
Page 52: ...4 2 GENERAL OPTIONS ...
Page 64: ......
Page 92: ......
Page 94: ...6 2 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Page 106: ...6 14 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Page 118: ...8 2 MAINTENANCE PERIPHERALS ...
Page 147: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 ...
Page 148: ...A1 2 APPENDIX 1 ...
Page 152: ......
Page 155: ...A3 1 APPENDIX 3 Appendix 3 Accesories A3 TITRATOR COMPONENTS A3 3 ...
Page 156: ......
Page 159: ...1 GENERAL APPLICATIONS BROCHURE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Page 160: ......
Page 161: ......
Page 163: ......
Page 165: ......
Page 167: ......
Page 169: ......
Page 171: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 177: ......
Page 179: ......
Page 181: ......
Page 183: ......
Page 185: ......
Page 187: ......
Page 189: ......
Page 192: ...2 TITRATION THEORY ...
Page 194: ...4 TITRATION THEORY ...
Page 206: ...16 TITRATION THEORY A B Figure 8 Potential glass electrode mV C ...