9-6
METHODS OPTIMIZATION
9.1.1.4 Start Mode
The
HI 903
can be set to either normal or cautious start mode. The cautious start feature is
designed to prevent the accidental over-titration of a sample with very low water content. In
cautious start mode, the HI 903 starts a titration using the minimum dose size specified by
the user rather than starting with half of the maximum dose size as with normal start mode.
9.1.1.5 Signal Averaging
The chosen value for the signal averaging setting determines how many readings the electronics
will average to produce a single data point on the titration curve. While higher values of 3 or
4 readings reduce the response time of the electrode, they also result in a ‘smoother’ titration
curve which may result in a faster titration (single unstable readings may cause the dose size
to be reduced).
9.1.1.6 Flow Rate
The flow rate setting specifies the volume of titrant delivered per minute. The default flow
rate should be used for the majority of titrations. In cases where the titrant is more viscous,
the flow rate can be reduced.
9.1.2
Termination Parameters
HI 903
provides a choice of three criteria by which a titration can be considered to have
reached an endpoint successfully.
9.1.2.1 Stability Time
When this termination criteria is selected, a titration is considered to have reached an endpoint
when the electrode potential stays below the specified endpoint potential for a period of time
called the stability time. Typical endpoint stability times range between 5 and 15 seconds.
In order for this criteria to successfully terminate a titration the stability time and the minimum
dose size must be set such that, at the end of a titration, the minimum dose size is large
enough to react all of the water leaking into the cell due to drift during the set stability time.
If the minimum dose volume is too small to compensate for the water introduced by the
drift, the titration will never be terminated.
9.1.2.2 Drift Stop Termination Criteria
Drift-based termination criteria, or Drift stop, terminates titrations based on the idea that at
the end of a titration, when all of the water due to the sample has been reacted, the titrator
should only be titrating the water seeping into the cell due to the background drift rate (see
section 6.3 for a detailed explanation of background drift).
Ideally, drift stop termination criteria would end a titration when a drift rate identical to that which
preceded the start of a titration is observed at the end of a titration. However, from a practical
standpoint the achievement of an identical drift rate results in very long titration times.
In order to shorten titration times while still taking advantage of the positive aspects of
Summary of Contents for HI 903
Page 1: ...1 QUICK START GUIDE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 0 ...
Page 4: ...4 QUICK START GUIDE ...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View ...
Page 18: ...18 QUICK START GUIDE QS 903 02 11 ...
Page 19: ...1 INSTRUCTION MANUAL HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Page 22: ......
Page 24: ...1 2 INTRODUCTION ...
Page 26: ...2 2 SETUP ...
Page 30: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View ...
Page 40: ...2 16 SETUP ...
Page 42: ...3 2 USER INTERFACE ...
Page 50: ......
Page 52: ...4 2 GENERAL OPTIONS ...
Page 64: ......
Page 92: ......
Page 94: ...6 2 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Page 106: ...6 14 TITRA TITRA TITRA TITRA TITRATION TION TION TION TION ...
Page 118: ...8 2 MAINTENANCE PERIPHERALS ...
Page 147: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 ...
Page 148: ...A1 2 APPENDIX 1 ...
Page 152: ......
Page 155: ...A3 1 APPENDIX 3 Appendix 3 Accesories A3 TITRATOR COMPONENTS A3 3 ...
Page 156: ......
Page 159: ...1 GENERAL APPLICATIONS BROCHURE HI 903 KARL FISCHER VOLUMETRIC TITRATOR Revision 1 00 ...
Page 160: ......
Page 161: ......
Page 163: ......
Page 165: ......
Page 167: ......
Page 169: ......
Page 171: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 177: ......
Page 179: ......
Page 181: ......
Page 183: ......
Page 185: ......
Page 187: ......
Page 189: ......
Page 192: ...2 TITRATION THEORY ...
Page 194: ...4 TITRATION THEORY ...
Page 206: ...16 TITRATION THEORY A B Figure 8 Potential glass electrode mV C ...