8. Assembly and Operation
Operation
1.
After assembly, install the belt and adjust the tension on the Tension pulley and spin the
Motor Pulley by hand to check.
2.
In the beginning, there is some vibration but as speed increases, it becomes more stable
(reaching 15,000 rpm/min in 1 minute).
If there are big noise and vibration, please stop the machine immediately and check it
out. Most of the case, it caused by wrong assembly.
3.
Only when there is no problem even if reaches setting rpm, power on the diaphragm
pump and start sample separation (But if there is no controller, rpm value is not
displayed, input the samples after 5 min).
4.
Automatic stop after setting time is finished (Because machine should not be stopped
during sample feeding, stop the sample feeding first and stop the motor)
5.
With stop the machine, because residual samples should be discharged through drain
on side, please install the Drip port.
Disassembly
Follow reverse order against assembly order