7.
Data on the Inverter and Diaphragm pump
<P/G Change when particular operating>
1.
In the input details of INVERTER SETTING
2.
No. 08 : Change b1-01=2 to b1-01=0
3.
No. 10 : Change b1-02=2 to b1-02=0
4.
No. 76 : Touch DATA/ENTER key
Example) In case the RPM is 15,000, press △, ▽,
▷
key and input 3480
The details of input values are as follows
RPM10,000 = 2,320RPM
RPM11,000 = 2,550RPM
RPM12,000 = 2,780RPM
RPM13,000 = 3,010RPM
RPM14,000 = 3,250RPM
RPM15,000 = 3,480RPM
RPM16,000 = 3,710RPM
Touch DATA/ENTER key
Turn the S/W off and turn the S/W on again after finishing input
And then, starting RUN and STOP