background image

 

 

Operating Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

J1250

 

Continuous Tubular Centrifuge 

Summary of Contents for J1250

Page 1: ...Operating Manual J1250 Continuous Tubular Centrifuge...

Page 2: ...opyright 2018 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contact Hanil Scientific Inc or place of purchase info ihanil com techsupport ihanil com Doc No OMJ1PDEN1801...

Page 3: ...hnical Specifications 1 2 Appearance 2 3 Structure 3 4 Installation 4 5 Precautions 6 6 LCD Display operation 8 7 Data on the Inverter and diaphragm pump 10 8 Assembly operation 16 9 Drag 18 10 Troubl...

Page 4: ...odel J1250 TUBULAR SEPARATOR Speed 15 000 RPM 15 800 xg Bowl Capacity 9 L Bowl Dimension inner 125 x 735 L mm Bowl Material Weight SUS 27 kg Time Control 99 hr 59 min 59 sec or Continuous 1 info ihani...

Page 5: ...2 Appearance Front View Side View Upper View info ihanil com techsupport ihanil com 2...

Page 6: ...3 Structure 3 info ihanil com techsupport ihanil com...

Page 7: ...uld be at least 60cm and empty the front space Electrical Wirings 1 Place the controller and control panel where easy to operate 2 After temporary wiring check the operation in low speed The rotation...

Page 8: ...ples in cushion tank when the samples to be transferred to other tank or transfer through other pipes This is for preventing the spouting 3 The optimal feeding pressure is between 0 1 0 5kg 4 The feed...

Page 9: ...to the machine It is particularly dangerous to touch the rotating part with gloves on 3 Please don t open bowl cover by when the machine stopped completely 4 Please check the system and parts regularl...

Page 10: ...ticular attention to the handling of the bowls and ensure that the threads are not scratched 7 Please use only plastic hammer for assembly or disassembly and don t use metal hammers 7 info ihanil com...

Page 11: ...Set operating time Up to 99 hr 59 min 59 sec or Continuous PROG Save or recall a program PUMP1 2 Each pump can be controlled and the pump is activated after the set rpm has been reached DOOR Display w...

Page 12: ...the setting value touch the Run button Level Sensor optional It is used to prevent overflow of the sample A beep sounds when 80 of the used sample container reaches The pump stops when it reaches 90 9...

Page 13: ...ontrol time for the stopping 10 6 C1 01 Acceleration time 400 7 C1 02 Deceleration time 600 8 E1 01 Input voltage setting 380 9 E1 04 Max voltage frequency 160 10 E1 05 Max voltage 380 11 E2 01 Motor...

Page 14: ...al time C1 02 OV Excess voltage Short deceleration time Increase the deceleration time C1 02 UV1 Low voltage High fluctuation voltage Use power transformer OL1 Motor overload Motor overload Increase t...

Page 15: ...er RPM setting Press ENTER Select parameters using arrow keys Setting all the parameters Press ENTER and press ESC to return to setting mode RPM Inverter Set Data RPM 10 000 2320 11 000 2550 12 000 27...

Page 16: ...nput 15000 on the RPM and press Enter 5 Time Example Input hour minute second and press Enter 1 Touch the Current Value 2 Input 4 32 on the Motor percent and press Enter Touch the lower part of the bo...

Page 17: ...erter RPM Inverter Input RPM Number RPM Set 102 67 4 32 6 2814 15000 3472 0 15000 Touch X after checking the above table and then start operating When you operate next time set up the RPM and Time Set...

Page 18: ...TA ENTER key Example In case the RPM is 15 000 press key and input 3480 The details of input values are as follows RPM10 000 2 320RPM RPM11 000 2 550RPM RPM12 000 2 780RPM RPM13 000 3 010RPM RPM14 000...

Page 19: ...ss down the spindle caution Part No 8 and 9 should be located in straight line 6 After confirm the clutch on no 4 connect the spindle and bowl using hammer and wrench 7 Press down the spindle sleeve c...

Page 20: ...by wrong assembly 3 Only when there is no problem even if reaches setting rpm power on the diaphragm pump and start sample separation But if there is no controller rpm value is not displayed input th...

Page 21: ...g from the frame by hand Drag can be removed on vertical direction 2 Please remove the drag shell together with spring using the drag shell wrench 3 The ferrule cap can be removed by hand If it doesn...

Page 22: ...it 3 If abrasion on the drag bushing is more than 1 6mm please twist half cycle and reuse it One bushing can be used twice Most of cases on the vibration caused by faulty bushing Please check dust or...

Page 23: ...9 Drag Drag Assembly Drawing Parts info ihanil com techsupport ihanil com 20...

Page 24: ...ushing 1 Is the balancing of the machine is correct Unbalancing creates big damages on the machine 2 Is the tension on all of the 6 bullet catches same Damages on the bowl boss sleeve Unbalance happen...

Page 25: ...ny abrasion on the drag bushing If there is abrasion more than 1mm it creates vibration on the bowl 7 Is the tension on the bullet catch and drag spring is got weakened Or is there any remaining oil o...

Page 26: ...spring please replace it with new 3 Is there any problem on the feed pipe line Damages on the tension pulley bearing 1 Was there any leakage on the tension pulley gasket If so please replace the ball...

Page 27: ...it damages on the rubber coupling Adjustment of belt tension depend on rotation direction 1 The rotation direction of motor is clockwise Please confirm the direction before connect belt For belt conne...

Page 28: ...lbow left and right 3 The adjustment of distortion on the belt a The adjustment of distortion depends on the horizontal angle of the tension pulley The contact surface between the belt and the tension...

Page 29: ...MEMO...

Page 30: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of Korea Tech Support 82 2 3452 8966 l techsupport ihanil com www ihanil com...

Reviews: