1 - 12
Quick Check® 600/800 Series User’s Guide
3.c To clear (delete) the records stored in the memory, press
↓
and
→
to
place the cursor on the asterisk next to the term
Clear
and then
press
SELECT
. The stored results and the result from the last scan
are deleted from the Quick Check 600/800 Verifier’s memory.
♦
To turn off the unit:
1. Press and hold down
POWER
for about two seconds until the unit beeps
once and clears the display.
OR
Let the unit turn itself off.
If the unit is connected to a charger, it turns off after ten
minutes of inactivity. If the unit is not connected to a charger, it turns off after
one minute of inactivity.
Note: If the unit turned itself off and you want to review the results of the last
scan, press
POWER
to turn on the unit and then press
←
. The results
from the last scan appear.
Calibrating the Quick Check 600/800 Verifier
Even though the verifier is calibrated at the factory, you should calibrate it before
its first use to ensure its accuracy. For continued accuracy, periodically calibrate
the unit during normal operation. You calibrate the unit using a reflectance page
that is included in the pack of accessories. Each page has a serial number on it.
For accurate calibration and data collection, you scan the reflectance page ten
times and the Quick Check 600/800 Verifier performs a consistency test. If it
passes, the unit is calibrated.
If, during the measurement of a symbol, a scan reflectance value of
XX
or
— —
appears on the display or on a printout, an out-of- range reading was recorded.
Recalibrate the unit and then rescan the symbol.
The reflectance page is a precision standard made of photographic materials. If
it becomes physically damaged or dirty, replace it to insure proper calibration. To
order a new reflectance page, contact your sales representative.
♦
To calibrate the Quick Check 600/800 Verifier:
1. Using the proper scan technique, scan the reflectance symbol from the
accessory pack. You will hear a long beep followed by two short beeps after
the first good scan, and the LCD displays
REFLECTANCE CALIBRATION
-9- More Scans!
.
2. Continue to scan the reflectance page for a total of ten good scans. After
each good scan, a single beep sounds and the display indicates how many
good scans you need to obtain. After the tenth good scan, three beeps
sound and the message
Now Lift the Scanner Up off the
Target
appears. When you lift the scanner up, the opening display
appears. To scan, set the wand scanner on a symbol’s quiet zone.
Summary of Contents for Quick Check 600 Series
Page 1: ... User s Guide Quick Check 600 800 Series Bar Code Verifiers ...
Page 22: ...1 16 Quick Check 600 800 Series User s Guide ...
Page 44: ...3 20 Quick Check 600 800 Series User s Guide ...
Page 47: ...Quick Check 600 800 Series User s Guide 4 3 Determining the ISO IEC Symbol Grade ...
Page 48: ...4 4 Quick Check 600 800 Series User s Guide Determining the ISO IEC Symbol Grade continued ...
Page 74: ...5 12 Quick Check 600 800 Series User s Guide ...
Page 88: ...7 6 Quick Check 600 800 Series User s Guide ...
Page 103: ...Quick Check 600 800 Series User s Guide 10 1 10 Flow Charts ...
Page 104: ...10 2 Quick Check 600 800 Series User s Guide ...
Page 105: ...Quick Check 600 800 Series User s Guide 10 3 ...
Page 106: ...10 4 Quick Check 600 800 Series User s Guide ...
Page 107: ...Quick Check 600 800 Series User s Guide 10 5 ...
Page 108: ...10 6 Quick Check 600 800 Series User s Guide ...
Page 109: ...Quick Check 600 800 Series User s Guide 11 1 11 SSTR Codes ...
Page 110: ...11 2 Quick Check 600 800 Series User s Guide ...
Page 111: ...Quick Check 600 800 Series User s Guide 12 1 12 Special Codes ...
Page 112: ...12 2 Quick Check 600 800 Series User s Guide ...
Page 114: ...13 2 Quick Check 600 800 Series User s Guide ...
Page 119: ......