Quick Check® 600/800 Series User’s Guide
4 - 13
Note: Certain high reflectance materials may always read “XX” or
“— —.” You should repeat the calibration.
The ratio of Rd/RI (Rmin/Rmax) must be less than 50% to pass this test.
Alternately, you can consider Rd < 50% RI. RI may be rated LO!, but it is never
rated HI!, while Rd may be rated HI! but is never rated LO!.
Symbol Contrast, Minimum and Maximum Reflectance
Symbol contrast, SymContr, is a graded measure of the difference between the
highest reflectance value R (max) and the lowest reflectance value, R (min),
obtained during a scan. SC = R(max)-R(min). Symbol contrast is rated A if >
70%, B if > 55%, C if > 40%, D if > 20% and
F if < 20%.
Edge Contrast
Edge Contrast minimum is the smallest edge contrast or “edge transition” in a
scan reflectance profile. The edge contrast is the difference between the
reflectance of an adjacent bar and space. EC (min) is the minimum value of the
edge contrast obtained during a scan. The criteria that edge contrast must be >
15% must be met. The EC (min) is rated A if > than 15% or F < 15%.
Modulation
Modulation is a ratio of the edge contrast minimum value divided by the symbol
contrast (EC (min)/SC). Modulation is rated A if > .70,
B > .60, C if > .50, D if > .40 and F if < .40.
Defects
Defects is the graded measure of the maximum element reflectance non
uniformity (noise within any element due to spots, voids, etc.), normalized to
symbol contrast. High readings are caused by print defects found:
•
in the symbol
•
adjacent to the quiet zone
•
at the transition of the label’s edge when a label which may have a marginal
or inadequate quiet zone is laid on top of another material.
Defects are rated A if < .15, B if < .20, C if < .25, D if < .30 and
F if > 30.
♦
To review the reflectance parameters:
1. Scan a symbol.
2. If the cursor is not next to Ref, use the arrows to place it there.
The first two reflectance parameters appear in the results line.
3. To view other reflectance parameters, press
↑
or
↓
.
The names of the reflectance parameters and their results scroll.
4. To review other types of data (dimensional or format), press
→
or
←
.
OR
Summary of Contents for Quick Check 600 Series
Page 1: ... User s Guide Quick Check 600 800 Series Bar Code Verifiers ...
Page 22: ...1 16 Quick Check 600 800 Series User s Guide ...
Page 44: ...3 20 Quick Check 600 800 Series User s Guide ...
Page 47: ...Quick Check 600 800 Series User s Guide 4 3 Determining the ISO IEC Symbol Grade ...
Page 48: ...4 4 Quick Check 600 800 Series User s Guide Determining the ISO IEC Symbol Grade continued ...
Page 74: ...5 12 Quick Check 600 800 Series User s Guide ...
Page 88: ...7 6 Quick Check 600 800 Series User s Guide ...
Page 103: ...Quick Check 600 800 Series User s Guide 10 1 10 Flow Charts ...
Page 104: ...10 2 Quick Check 600 800 Series User s Guide ...
Page 105: ...Quick Check 600 800 Series User s Guide 10 3 ...
Page 106: ...10 4 Quick Check 600 800 Series User s Guide ...
Page 107: ...Quick Check 600 800 Series User s Guide 10 5 ...
Page 108: ...10 6 Quick Check 600 800 Series User s Guide ...
Page 109: ...Quick Check 600 800 Series User s Guide 11 1 11 SSTR Codes ...
Page 110: ...11 2 Quick Check 600 800 Series User s Guide ...
Page 111: ...Quick Check 600 800 Series User s Guide 12 1 12 Special Codes ...
Page 112: ...12 2 Quick Check 600 800 Series User s Guide ...
Page 114: ...13 2 Quick Check 600 800 Series User s Guide ...
Page 119: ......