57
10.
Adjust the yaw adjustment on
the laser base so that LP#2 is
tilted until both Target #2 and #3
show the same readings. LP#2 is
now parallel to the offset
centerline.
Note:
The readings do not have to
be zero, but just the same number
and same sign.
11.
To measure one of the rolls for
parallelism, place Target #1
horizontally on the roll closest to
the laser. Adjust the post so the
laser plane is near the center of
the target. Move the magnetic
base up/down the roll until the
bubble is level. Zero Target #1
in Read15.
12.
Move Target #1 to the far end of
the roll. The resulting value is a
measure of the parallelism of
this roll relative to the reference
roll. In this case, a
plus (
+)
reading means the far end of the
roll is pointing to the right by
.025 in. To align it, adjust the
roll until the readout shows
zero, which means the roll is
aligned and parallel to the
reference roll.
Summary of Contents for STEALTH Series
Page 1: ......
Page 69: ...65 Appendix A Equipment Drawings L 733 Precision Geometry Laser...
Page 70: ...66 L 743 Ultra Precision Geometry Laser...
Page 71: ...67 A 1519 A 1520 Universal Wireless Targets...
Page 72: ...68...
Page 73: ...69...
Page 74: ...70...
Page 75: ...71...
Page 76: ...72...