33
Aligning the Laser to the X-Axis Travel
1.
Mount Target #4 on the ram, move the
spindle and adjust the target height.
2.
Power on the laser and zero the target
in the Read9 PDA. Move the column
down to the end of the X-Axis travel
and rotate the target head to point
towards the laser.
3.
Turn the Yaw-Axis adjustment knob to
center the laser to zero on the target.
4.
Return the column to its starting
position and rotate the target toward
the laser. Zero the target.
5.
Send the column back to the far end of
travel and tweak the Yaw-Axis
adjustment knob to zero the laser.
6.
When the target reads zero at both
ends, the laser is aligned to the X-Axis.