33
Settings
Affected Mode
Parameter
DESCRIPTION
Range/Units
Default Value
Increment
IDL Mode
SECONDARY THERMOCOUPLE
Disable optional Thermocouple Probe
ENABLE/ DISABLE
DISABLE
N/A
INITIAL TIME
Ramp-up interval to INITIAL TEMP set point
15 – 120 s
30 s
1 s
INITIAL TEMP
Initial target for temperature ramping.
40° – 80°C
65°C
1°C
RAMP RATE
Rate at which temperature increases from the
INITIAL TEMP to the SET TEMP
0.5° – 5°C/min
2°C/min
0.1°C/min
POWER LIMIT
Maximum power available for temperature control
1 – 50 W
50 W
1 W
7.8 Special Event States
7.8.1 Recoverable
Errors
Figure 7-30 Recoverable Error Pop-Up Display
7.8.1.1 Display
• Yellow “error” indicator will display on screen.
• Error text will display in the center of the screen within the Status window.
7.8.1.2 Settings
• In all modes except Mult-Cool, a 10-second time-out returns device to previous READY mode. If the probe and cable have been detached the Generator returns to the
STANDBY state.
• In Multi-Cool, a 60-second time-out returns device to previous READY mode. If the probe and cable have been detached the Generator returns to STANDBY state.
• Recoverable Error: measurement values are shown, and settings are greyed out.
7.8.2 Non-Recoverable
Faults
Figure 7-31 Non-Recoverable Fault Pop-Up Display