7.3 READY
States
• VOLTAGE STIMULATION Mode – READY state is the default upon connection of a
HALYARD*
Pain Management Thermocouple probe and cable.
Figure 7-5 VOLTAGE STIMULATION – READY Display
Figure 7-6 MULTI-RF VOLTAGE STIMULATION – READY Display
Displays and Indicators
PARAMETER
STATUS
DESCRIPTION
RF/Stimulus Output
OFF
RF power is not delivered during the READY state. When Output ON/OFF is pressed, RF power is delivered and the
Generator automatically transitions to the ON state.
VOLTAGE STIMULATION – During the ON state, Voltage must be greater than 0.00 V for Stimulus pulses to be
delivered.
CURRENT STIMULATION – Stimulus pulses are not delivered during the READY state. During the ON state, Current
must be greater than 0.00 mA for Stimulus pulses to be delivered.
TRANSDISCAL AND COOLED RF TREATMENT – When Output ON/OFF is pressed; the Generator will enter the PRE-
TREATMENT COOLING state.
FAULT Indicator
OFF
(The FAULT indicator is used only in the POST and SYSTEM FAULT states.)
RF/Stimulus Output
ON/OFF Indicator
OFF
The indicator is not illuminated during the READY state, as there is no Stimulus/RF power delivery.
RF Audio Output
None
RF audio output is not generated during the READY state.