10
8. FOOTSWITCH
Connection:
This pneumatic barb connects to the hose of the pneumatic footswitch. Like the OUTPUT ON/OFF switch, the FOOTSWITCH controls
stimulus output in the STIMULATION modes and controls RF output in the LESION, COOLED RF, TRANSDISCAL, RFA and IDL modes. The action of the FOOTSWITCH
differs from the OUTPUT ON/OFF switch: the FOOTSWITCH must be held down to continue delivery of stimulus pulses or RF energy.
9. RS-232
Connection:
This is an isolated connection to a standard 9-pin COM port. It is provided for passive data acquisition, and cannot be used to control the
system.
10. RJ45
Connector:
Connection to be used by authorized service personnel only.
11. Pump Module Interface Connector:
For attachment of authorized cooling pump unit only.
6 Factory Set Defaults
Factory Set Defaults
Mode
Settings
Adjustment Range
“Factory Set” Value
VOLTAGE
STIMULATION
STIM RATE
1-Shot, 2, 5, 10, 20, 50, 75, 100, 150, 180, 200 Hz
2 Hz
PULSE DUR
0.1, 0.2, 0.5, 1.0 ms
1.0 ms
VOLTAGE
0.0 –10.0 V
0.0 V
CURRENT
STIMULATION
STIM RATE
1-Shot, 2, 5, 10, 20, 50, 75, 100, 150, 180, 200 Hz
2 Hz
PULSE DUR
0.1, 0.2, 0.5, 1.0 ms
1.0 ms
CURRENT
0.0 –10.0 mA
0.0 mA
AUTO TEMP LESION
SET TEMP
38° – 95°C
80°C
TIME
15 – 600 s
75 s
MULTI-RF (TRUE-TX)
MAX RAMP TIME EXTENSION
DISABLE, 5–120 s, with 5 s interval
30 s
MANUAL POWER
LESION
TEMP LIMIT
38° – 95°C
80°C
TIME
15 – 600 s
75 s
SET POWER
0 – 50 W
0 W
AUTO PULSED
LESION
SET TEMP
38° – 95°C
42°C
TIME
10 – 900 s
120 s
PULSE DUR
10 – 100 ms, with 10 ms interval
20 ms
PULSE RATE
1 – 10 Hz
2 Hz
MANUAL PULSED
LESION
WARNING TEMP
38° – 95°C
42°C
TIME
10 – 900 s
120 s
PULSE DUR
10 – 100 ms, with 10 ms interval
20 ms
PULSE RATE
1 – 10 Hz
2 Hz
COOLED RF
SET TEMP
30° – 90°C
60°C
TIME
15 s–30 min
2 min 30 sec
MULTI-COOLED RF
MAX RAMP TIME EXTENSION
DISABLE, 5–180 s, with 5 s interval
60 s
SET TEMP
38° – 95°C
60°C
IMPEDANCE CUTOFF MULTIPLIER
DISABLE, 2.0 – 4.0, with 0.1 interval
2.0
TIME (1 Probe)
15 s – 30 min with 15 s interval
2 min 30 sec
TIME (2 Probes)
15 s – 30 min with 15 s interval
2 min 30 sec
TIME (3 – 4 Probes)
15 s – 30 min with 15 s interval
2 min 45 sec
TRANSDISCAL
SET TEMP
30° – 90°C
BIPOLAR 50°C
MONOPOLAR 60°C
TIME
1 – 30 min
BIPOLAR 15 min
MONOPOLAR 2 min 30 sec
RFA
SET TEMP
38° – 95°C
55°C
TIME
1 – 30 min
10 min
IDL
SET TEMP
40° – 95°C
90°C
TIME
30 s – 20 min
16 min 30 sec
Avanos Exhibit 2069 Page 10