7.7.5 ADVANCED SETTINGS Mode – COOLED RF SETTINGS State
The following screen appears when COOLED RF SETTINGS is selected from the ADVANCED SETTINGS Mode:
• This screen allows adjustment of the RAMP RATE used by the COOLED RF mode to attain the SET TEMPERATURE.
• The option to use POST TREATMENT COOLING after the active treatment has ended can be accessed by toggling the ENABLE/DISABLE Post Treatment Cooling control.
• The option to use one or two probes is accessible by adjusting NUMBER OF PROBES.
• The maximum amount of power available can be adjusted by rotating the dial to change SET POWER LIMIT.
Figure 7-26 ADVANCED SETTINGS Mode – COOLED RF SETTINGS Display
Settings
Affected Mode
Parameter
DESCRIPTION
Range/Units
Default Value
Increment
Cooled RF Auto
Temp Mode
RAMP RATE
Rate at which temperature increases from the INITIAL TEMP to
the SET TEMP.
0.5° – 5°C/min
80°C/Min
0.1°C/Min
5° – 10°C/min
5°C/Min
10° – 200°C/min
10°C/Min
POST TREATMENT
COOLING
Select to enable or disable the option to go into POST TREATMENT
COOLING after treatment.
ENABLE/DISABLE
DISABLE
N/A
Stimulation Mode
Cooled RF Auto
Temp Mode
NUMBER OF
PROBES
Number of Cooled RF Probes used.
1, 2
1
Cooled RF Auto
Temp Mode
POWER LIMIT
Maximum power available for temperature control.
1–50 W
25 W
1 W