STATE
PARAMETER
Measurement Window
Available Modes
Available Settings
Status
Description
Status
Description
Status
Description
COOLED RF
TREATMENT
ON
The temperature of the Cooled RF probe(s)
are displayed in the COOLED RF TEMP box.
Impedance measurement between the
Cooled RF probes are displayed. Time and
Power are not displayed (dashed lines
replace values).
VOLTAGE
STIMULATION
CURRENT
STIMULATION
SAVE
SETTINGS
While in the READY state,
Placement is available. When
Placement mode is selected,
the Generator will enter the
READY state of that mode.
The present SETTINGS can be
saved as defaults by holding
down the SAVE SETTINGS key.
SET TEMP
Range
38° – 95°C
Increment 1°C
Default**
60°C ***
TIME
Range
†
15 s – 30 min
Increment
15 s (between 15 s and
5 min) 30 s (between 5
min and 30 min)
Default**
2 min 30 sec
TRANSDISCAL
TREATMENT
The temperature of the
TRANSDISCAL*
probe(s) is displayed in the TRANSDISCAL
TEMP box. The temperature of Peripheral
Disc Temperature is displayed in the
PERIPHERAL DISC TEMP box.
Impedance measurement between
TRANSDISCAL*
probes are displayed. Time
and Power are not displayed (dashed lines
replace values).
VOLTAGE
STIMULATION
CURRENT
STIMULATION
PLACEMENT
MODE
SAVE
SETTINGS
SET TEMP
Range
30° – 90°C
Increment 1°C
Default**
BIPOLAR 50°C
MONOPOLAR 60°C
TIME
Range
†
1 – 30 min
Increment 30 s
Default**
BIPOLAR 15 min
MONOPOLAR
2 min 30 sec
RFA
TREATMENT
MODE
If enabled, Temperature of RFA and
Thermocouple Probe are displayed.
Impedance measurement for RFA is
displayed. Time and Power are not displayed
(dashed lines replace values). If the
Thermocouple Probe is disabled via RFA
SETTINGS in the ADVANCED SETTINGS state,
that measurement is greyed out and the
value replaced with dashed lines.
PLACEMENT
MODE
SAVE
SETTINGS
SET TEMP
Range
38° – 95°C
Increment 1°C
Default**
55°C
TIME
Range
1 – 30 min
Increment 30 s
Default**
10 min
IDL MODE
Temperature of IDL and Thermocouple
Probe are displayed. Impedance
measurement for IDL is displayed. Time
and Power are not displayed (dashed lines
replace values). If the Thermocouple Probe
is disabled via IDL SETTINGS state, accessible
from the ADVANCED SETTINGS mode, that
measurement is greyed out and the value
replaced with dashed lines.
SAVE
SETTINGS
While in the READY state,
other operating modes are
not available.
The present SETTINGS can be
saved as defaults by holding
down the SAVE SETTINGS key.
SET TEMP
Range
40° – 95°C
Increment 1°C
Default**
90°C
TIME
Range
30 s – 20 min
Increment 00:30 s
Default**
16:30 min
** Defaults are factory-set, and are overwritten when the SAVE SETTINGS function is used.
*** The “Cooled RF Set Temp” (Default Setting T = 60°C) displayed on the COOLIEF* RF Generator refers to the cooled electrode temperature and does not reflect the immediate
surrounding tissue temperature. The heat generated from the radiofrequency energy produces thermal energy with average maximum tissue temperatures greater than 80°C.
†
Time Range set by user does not include factory-set PRE-TREATMENT COOLING Time (45 seconds) or POST-TREATMENT COOLING Time (10 seconds).