![Halyard CO LIEF PMG-Basic User Manual Download Page 16](http://html1.mh-extra.com/html/halyard/co-lief-pmg-basic/co-lief-pmg-basic_user-manual_3379182016.webp)
16
STATE
PARAMETER
Measurement Window
Available States
Available Settings
Status
Description
Status
Description
Status
Description
AUTO TEMP
LESION
ON
Temperature
and Impedance
measurements are
displayed while Time
and Power are not
displayed (dashed
lines replace values).
VOLTAGE
STIMULATION
CURRENT
STIMULATION
AUTO TEMP
MANUAL
POWER
AUTO PULSED
MANUAL
PULSED
SAVE
SETTINGS
While in the READY state,
other operating modes for
HALYARD*
Pain Management
Thermocouple probes are
available. When a different
mode is selected, the Generator
will enter the READY state of
the new mode.
The present SETTINGS can be
saved as defaults by holding
down the SAVE SETTINGS key.
Multi-Cooled RF only supports
AUTO TEMP LESION mode.
SET TEMP
Range
38° – 95°C
Increment 1°C
Default**
80°C
TIME
Range
15 – 600 s
Increment 1 s
Default**
75 s
MANUAL
POWER
LESION
TEMP
LIMIT
Range
38° – 95°C
Increment 1°C
Default**
80°C
TIME
Range
15 – 600 s
Increment 1 s
Default**
75 s
POWER
LIMIT
Range
0 – 50 W
Increment N/A
Default
0 W
(Cannot be overwritten by SAVE SETTINGS function)
AUTO
PULSED
LESION
SET TEMP
Range
38° – 95°C
Increment 1°C
Default**
42°C
TIME
Range
10 – 900 s
Increment 1 s
Default**
120 s
PULSE
DUR
Range
10 –100 ms
Increment 10 ms
Default**
20 ms
PULSE
RATE
Range
1 –10 Hz
Increment 1 Hz
Default**
2 Hz
MANUAL
PULSED
LESION
WARNING
TEMP
Range
38° – 95°C
Increment 1°C
Default**
42°C
TIME
Range
10 – 900 s
Increment 1 s
Default**
120 s
PULSE
DUR
Range
10 –100 ms
Increment 10 ms
Default**
20 ms
PULSE
RATE
Range
1 – 10 Hz
Increment 1 Hz
Default**
2 Hz
Note:
Multi-RF Manual Power and Manual Pulsed modes support 1 probe connected to the Multi-RF module.
** Defaults are factory-set, and are overwritten when the SAVE SETTINGS function is used.
Avanos Exhibit 2069 Page 16
Medtronic, Inc. v. Avanos Medical Sales, LLC
Case IPR2020-0089