28
7.7.4.1 ADVANCED SETTINGS Mode – MULTI-COOLED RF1 SETTINGS State
The following screen appears when MULTI-COOLED RF1 SETTINGS is selected from the ADVANCED SETTINGS Mode:
• This screen allows adjustment of POWER LIMIT in AUTO TEMP mode.
• This screen allows adjustment of the initial RAMP RATE taken by the AUTO TEMP mode to attain the SET TEMPERATURE.
• Changing RAMP RATE affects the portion of the total TIME spent at the SET TEMPERATURE.
• Raising the RAMP RATE causes the Generator to attempt to reach the SET TEMPERATURE faster.
• MAX RAMP TIME EXTENSION is the maximum allowable duration for extension of the ramp time in Multi-RF and Multi-Cooled RF Auto Temp Lesion mode, to maintain a
consistent Lesion Time. If all probes have not reached Set Temperature within the sum of this time and the specified Ramp Time, an error notification will be generated. This
feature can be turned off by changing the setting to DISABLE.
Figure 7-25a ADVANCED SETTINGS Mode — MULTI-COOLED RF1 SETTINGS Display
Settings
Affected Mode
Parameter
DESCRIPTION
Range/Units
Default Value
Increment
AUTO TEMP MODE
POWER LIMIT
Maximum allowable power for temperature control. 1 – 50 W
50 W
1 W
AUTO TEMP MODE
1 Probe
RAMP RATE
Ramp-up interval to temperature set point.
0.5° – 5°C/min
80°C/min
0.1°C/min
5° – 10°C/min
5°C/min
10°–200°C/min
10°C/min
AUTO TEMP MODE
2 Probes
0.5° – 5°C/min
60°C/min
0.1°C/min
5° – 10°C/min
5°C/min
10°–200°C/min
10°C/min
AUTO TEMP MODE
3 – 4 Probes
0.5° – 5°C/min
40°C/min
0.1°C/min
5° – 10°C/min
5°C/min
10°–200°C/min
10°C/min
MULTI-COOLED RF
(TRUE-TX)
MAX RAMP TIME
EXTENSION
Maximum allowable duration for extension of ramp
time. If all probes have not reached Set Temperature
within this time, an error notification will be
generated.
DISABLE,
5 – 120 s
30 s
5 s