background image

Reduce the time of keeping door open when storing or taking food each time, to prevent great 
fluctuation of the temperature and humidity inside the refrigerator. 

The refrigerator will see a sharp temperature increase in a short time when the door is opened, 
which is normal, and the temperature will resume 1h after the door is closed. 

The refrigerator shall operate with ambient temperature as 18 ~ 38°C
 (temperature of HXC-1308/1308B: 16 ~ 32°C) and humidity less than 85%; condensation will 
occur on the glass door in high temperature and high humidity condition, which is normal and will
not affect the storage temperature inside the refrigerator; in the event of condensation, please
improve ventilation conditions as soon as possible and reduce ambient temperature simultaneously. 

The temperature display value of refrigerator refers to the temperature value at the temperature
sensing probe inside the refrigerator. Though the temperature displayed is different from the 
actual one at the center of refrigerator, it will approach to the real temperature gradually. 

The refrigerator battery may be low when the refrigerator is restarted after it has been 
out-of-service for a long time; make sure the battery switch is set at “ON” and charge the storage 
battery, which will be full after continuous running of refrigerator about a week. 

Make sure the temperature inside the refrigerator has reached the set value and put articles in the
refrigerator by batch. Volumes accounted by articles shall not be greater than 1/3 of refrigerator to 
prevent rise of temperature. 

Clean the freezer with neutral detergent diluted. Brushes, acids, gasoline, soap powders, polishing 
powders or hot water are prohibited to clean refrigerator as these materials may damage the 
painting surface and plastic & rubber components and parts. Never wipe the plastic & rubber 
components and parts with volatile solvent such as gasoline. 

Shut off the power as well as the battery switch if the refrigerator will be out-of-service for a long 
time. 

39

6

Precautions for Use

Specification

HXC-608

115V/60Hz

Blood Bank Refrigerator 

External dimensions (W X D X H) 

mm

 

Internal dimensions (W X D X H) 

mm

 

Effective volume 

Door 

Insulating materials 

Compressor 

Shelf / Drawer

Refrigeration mode 

Outer plate / inner plate 
Condenser / evaporator 

Refrigerant 

Headlamp 

Net weight 

Temperature of temperature 

measurement box 

Rated power supply 

Rated power / current 

Alarm system 

Rechargeable battery 

780x840x1945 

680x640x1400 

608L 

Electrical heating glass door 

Hard polyurethane foam (fluorine-free) 

Totally enclosed imported compressor 

Forced Wind Cooling Circulation 

Spraying steel plate / stainless steel plate 

Fin / fin 

Computer control system 

R134a 420g  

LED   3W

204kg

4

±1

115V/60Hz

560W/7.5A

High-low temperature alarm, sensor fault alarm, power failure 

alarm,  low battery alarm and door opening alarm 

DC 12 V rechargeable battery 

Name 
Model 

Duration time of power failure alarm

48 hours (when battery is fully charged) 

I

Temperature controller 

USB function / temperature recorder 

Standard / optional 

Y

Electric shock resistance type 

Power connect style

Climate Type

Foaming Agent

Noise Level

4

43dB(A)

CP/IP

Note: Climate Type 4 means the temperature is +30

,the relative humidity is 55%.Technical 

information may vary from the appliance that you have just purchased due to technical improvements.

The loading of each shelf

Max:30kg

6 shelves / 24 baskets 

Do not dispose of electrical appliances as unsorted municipal waste, use separate collection 
facilities.Contact you local government for information regarding the collection systems available.
If electrical appliances are disposed of in landfills or dumps, hazardous substances can leak into 
the groundwater and get into the food chain, damaging your health and well-being.When replacing 
old appliances with new ones, the retailer is legally obligated to take back your old appliance for 
disposals at least free of charge.

Meaning of crossed –out wheeled dustbin:

Summary of Contents for HXC-158B

Page 1: ...echnology Development Zone Qingdao 266510 P R China Web www haiermedical com Version 1st 2016 Dedicated code 0270501029 V13026 Read the Operation Manual carefully before using your appliance Keep the...

Page 2: ......

Page 3: ...ent from the figure in the manual due to product improvement The manual may be subject to changes without further notice Computer panel control digital display of temperature temperature display accur...

Page 4: ...ety Precautions Precautions for Use Product Installation Component Names Control Panel Application method Alarm Cleaning and Maintenance FAQ Refrigeration Schematic Circuit Diagram Specification 2 3 6...

Page 5: ...t to prevent personal injuries and refrigerator damage In case of flammable gas leakage such as coal gas shut off the valve of leaked gas open windows for ventilation and exhaust do not plug in or pul...

Page 6: ...gged into the socket to prevent fire Apply socket with ground wire to prevent electric shock If the socket fails to be grounded ground wires must be installed by technician Do not use the refrigerator...

Page 7: ...o children And the flammable foaming needs to be disposed by professional persons CP IP foaming materials are flammable need professional processing HXC 608B 6 Drawers HXC 608 Blood Bank Refrigerator...

Page 8: ...solvent such as gasoline Shut off the power as well as the battery switch if the refrigerator will be out of service for a long time 39 6 Precautions for Use Specification HXC 608 115V 60Hz Blood Ban...

Page 9: ...The refrigerator shall be grounded in a reliable manner If the socket is provided with grounding wire check if it is well grounded before operation If the socket is not provided with grounding wire a...

Page 10: ...ilure alarm 48 hours when battery is fully charged I Temperature controller USB function temperature recorder Optional optional Y Electric shock resistance type Power connect style Climate Type Foamin...

Page 11: ...control panel Internal fan Display panel Sensor Defrosting Control Lower temperature Upper temperature Alarm VCC GND Heating strip at the cabinet outlet Heating strip of center sill Door heating trans...

Page 12: ...the articles gradually into the refrigerator after complete examination of refrigerator 35 10 Circuit Diagram HXC 158 158B 358 358B 608 608B CN2 Cooling fan Compressor Defrosting heating strip CN2 CN...

Page 13: ...rature at one time If so the compressor may not stop for a long time with temperature reducing slowly which will reduce service life Articles must be put in batches and the temperature shall be reduce...

Page 14: ...ean data tested in the standard laboratory without noise in which the refrigerator without storage of articles is laid on the rubber pad and is steadily operated after closing the door to measure out...

Page 15: ...e for detailed information Purchasing of backup battery of the recorder Battery in the same specification and the same brand can be purchased in proper shops 9V DC In order to guarantee the reliabilit...

Page 16: ...pment and may lead to the occurrence of fault Compressor and other mechanical parts are in fully closed state and they do not need to be lubricated Clean the refrigerator Cleaning and Maintenance Comp...

Page 17: ...attery switch is on or off the remote alarm can be started The remote alarm port coordinates with the audible alarm on the refrigerator Thus the remote alarm can be stopped by pressing the key of buzz...

Page 18: ...or the refrigerator He is in charge of examining the running status of the refrigerator and making the record If the temperature inside the refrigerator is too high or too low or other abnormal phenom...

Page 19: ...ack to the normal display mode The original temperature set value 4 flickers Please do the next operation within 60s Otherwise the display panel will return to display TS SN Alarm items Buzzing Light...

Page 20: ...fter a period of Fd user parameters can be set the default value is 30 min the buzzer and remote alarm will be restarted 1Hz Continuously Buzzing 1Hz Continuously Buzzing Alarm light flashes Display b...

Page 21: ...til the pen returns to the paper At this moment the pen will pause at the most external temperature scaling line of the recording paper this scaling line does not have to be marked with temperature va...

Page 22: ...g the change of temperature in the box In order to use the disk recorder in a right and reasonable way please read the instruction bellow carefully Get familiar with your temperature recorder Adjustme...

Page 23: ...ction The computer panel has USB output function which can make use of U disk to export temperature and other test data Plug in U disk the machine begins to automatic identification and import data in...

Page 24: ...position can also be adjusted finely 3s later the scriber will slide to appointed scale according to the present perceived temperature then the calibration ends After retightening hold down bolt and...