background image

29

16

Temperature setting

To modify the temperature set value, for example, the original set value 4°C is to be modified as 
4.5°C, please operate according to the following steps: 
Temperature setting

HXC-158/158B/358/358B/608/608B

 

2

1

3

4

Button operation

Display

Press on “Sensor” and “Cal cancel” at the 
same time for more than 5 seconds

Click “Cal cancel” (increase 0.1°C/time) or
 “Sensor” (reduce 0.1°C/time)

Stop operation after the required 
temperature value is set as 4.5°C

Display the temperature in box

Original set value  4°C appears and flashes 
(please conduct next operation within 5s of 
the flashing, otherwise, the displayer will 
restore the temperature in the display box)

The flashing of displayed temperature value
 4.5°C stops after 5 seconds. Temperature 
setting value is stored in the system. The 
displayer redisplays the temperature in box 
if the setting works.

The original set value 4°C will change 
accordingly until the displayed value on the 
display screen reaches 4.5°C

When there is a power outage, the fully charged backup battery can enable the alarm function to 
keep working for 48 hours For the initial startup of the refrigerator or after a long time of power off, 
the battery must be fully charged and the refrigerator needs to be run for 2 days. 
Press key of “Alarm test”. Each pressing will enable the buzzer to make 3 times of sound with the 
frequency of continuous 1 Hz, meanwhile, the alarm indicator light flashes for 3 times. When it is 
not in the remote hardware alarming state, the remote alarm relay will be disconnected after having 
operated for 3 times. Then decide which action shall be taken according to whether alarm is 
needed. And this indicates that the alarm system is in normal conditions, or else the alarm system 
has fault. 
Someone appointed shall be responsible for the refrigerator. He is in charge of examining the
running status of the refrigerator and making the record. If the temperature inside the refrigerator 
is too high or too low or other abnormal phenomenon, he shall transfer the articles to other 
refrigerators which conform to the conditions till the fault has been settled and articles can be 
stored in that refrigerator again. 
When the battery has fault or the connecting wire has error, press “Alarm test” and “E5” is 
displayed in the display area. At the moment, examine whether the battery switch is open or closed 
and whether the connection is correct. Replace the battery if all of the above are correct (except for 
HXC-1308/1308B). 
When the switch of the rechargeable battery is not open or the there is less battery power, press 
the key of alarm test for one time, and “E6” is flashing and displaying in the displaying window for 
3 times (be applicable to HXC-1308/1308B) .

SN. Alarm items 

Buzzing

Light 

flashing 

5

Alarm of 

super high 

temperature 

Conditions of alarm start 

Display 

requirements 

Note 

Display the actual 
conditions 

High temperature controller 
detects the temperature inside 

the product is ≥ 12; at the time, 

both upper and lower sensors 
are bound to be detected with 
temperature excess. 

Alarm of simultaneous 
temperature excess 

Alarm of simultaneous 
temperature excess 

6

Remote 

alarm 

(1) Temperature excess in 
upper or lower sensor; 
(2) Fault in upper, lower, control, 
defrosting and alarming sensors; 
(3) AC220V/110V power outage . 
Enable remote alarm when any 
one of the above conditions is
 met. 

Accompanied 
with 
continuous 
buzzing with 
1 Hz 

Accord-
ing to 
different 
alarming 
situations, 
execute 
corresp-
onding to 
item 1, 2 
and 3. 

According to 
different alarming 
situations, execute 
corresponding to
item 1, 2 and 3. 

According to different 
alarming situations, 
execute corresponding 
to item 1, 2 and 3. 
Remote alarm has NO 
and NC functions. 
When the power is cut 
off outside, no matter 
whether the battery 
switch is on or off, the 
remote alarm can be 
started. 

Summary of Contents for HXC-158B

Page 1: ...echnology Development Zone Qingdao 266510 P R China Web www haiermedical com Version 1st 2016 Dedicated code 0270501029 V13026 Read the Operation Manual carefully before using your appliance Keep the...

Page 2: ......

Page 3: ...ent from the figure in the manual due to product improvement The manual may be subject to changes without further notice Computer panel control digital display of temperature temperature display accur...

Page 4: ...ety Precautions Precautions for Use Product Installation Component Names Control Panel Application method Alarm Cleaning and Maintenance FAQ Refrigeration Schematic Circuit Diagram Specification 2 3 6...

Page 5: ...t to prevent personal injuries and refrigerator damage In case of flammable gas leakage such as coal gas shut off the valve of leaked gas open windows for ventilation and exhaust do not plug in or pul...

Page 6: ...gged into the socket to prevent fire Apply socket with ground wire to prevent electric shock If the socket fails to be grounded ground wires must be installed by technician Do not use the refrigerator...

Page 7: ...o children And the flammable foaming needs to be disposed by professional persons CP IP foaming materials are flammable need professional processing HXC 608B 6 Drawers HXC 608 Blood Bank Refrigerator...

Page 8: ...solvent such as gasoline Shut off the power as well as the battery switch if the refrigerator will be out of service for a long time 39 6 Precautions for Use Specification HXC 608 115V 60Hz Blood Ban...

Page 9: ...The refrigerator shall be grounded in a reliable manner If the socket is provided with grounding wire check if it is well grounded before operation If the socket is not provided with grounding wire a...

Page 10: ...ilure alarm 48 hours when battery is fully charged I Temperature controller USB function temperature recorder Optional optional Y Electric shock resistance type Power connect style Climate Type Foamin...

Page 11: ...control panel Internal fan Display panel Sensor Defrosting Control Lower temperature Upper temperature Alarm VCC GND Heating strip at the cabinet outlet Heating strip of center sill Door heating trans...

Page 12: ...the articles gradually into the refrigerator after complete examination of refrigerator 35 10 Circuit Diagram HXC 158 158B 358 358B 608 608B CN2 Cooling fan Compressor Defrosting heating strip CN2 CN...

Page 13: ...rature at one time If so the compressor may not stop for a long time with temperature reducing slowly which will reduce service life Articles must be put in batches and the temperature shall be reduce...

Page 14: ...ean data tested in the standard laboratory without noise in which the refrigerator without storage of articles is laid on the rubber pad and is steadily operated after closing the door to measure out...

Page 15: ...e for detailed information Purchasing of backup battery of the recorder Battery in the same specification and the same brand can be purchased in proper shops 9V DC In order to guarantee the reliabilit...

Page 16: ...pment and may lead to the occurrence of fault Compressor and other mechanical parts are in fully closed state and they do not need to be lubricated Clean the refrigerator Cleaning and Maintenance Comp...

Page 17: ...attery switch is on or off the remote alarm can be started The remote alarm port coordinates with the audible alarm on the refrigerator Thus the remote alarm can be stopped by pressing the key of buzz...

Page 18: ...or the refrigerator He is in charge of examining the running status of the refrigerator and making the record If the temperature inside the refrigerator is too high or too low or other abnormal phenom...

Page 19: ...ack to the normal display mode The original temperature set value 4 flickers Please do the next operation within 60s Otherwise the display panel will return to display TS SN Alarm items Buzzing Light...

Page 20: ...fter a period of Fd user parameters can be set the default value is 30 min the buzzer and remote alarm will be restarted 1Hz Continuously Buzzing 1Hz Continuously Buzzing Alarm light flashes Display b...

Page 21: ...til the pen returns to the paper At this moment the pen will pause at the most external temperature scaling line of the recording paper this scaling line does not have to be marked with temperature va...

Page 22: ...g the change of temperature in the box In order to use the disk recorder in a right and reasonable way please read the instruction bellow carefully Get familiar with your temperature recorder Adjustme...

Page 23: ...ction The computer panel has USB output function which can make use of U disk to export temperature and other test data Plug in U disk the machine begins to automatic identification and import data in...

Page 24: ...position can also be adjusted finely 3s later the scriber will slide to appointed scale according to the present perceived temperature then the calibration ends After retightening hold down bolt and...

Reviews: