SN. Alarm items 

Buzzing

Light 

flashing 

1

2

3

Over 

temperature 

alarm (high 

and low 

temperature 

alarm) 

Fault of 
sensor 

Power 
failure 
alarm 

Conditions of alarm start 

When any temperature of upper 
and lower display temperature is 
no lower than ALH (the alarm 
temperature is of upper limit and 
the user parameter can be set), 
high temperature alarm starts; 
when any temperature of upper 
and lower display temperature is 
no higher than ALH (the alarm 
temperature is of lower limit and 
the user parameter can be set), 
lowtemperature alarm starts. 

Display 

requirements 

Note 

1Hz 

Continuously 

Buzzing 

Alarm 

light 

flashes 

Display actual 

temperature 

Start remote alarm, 
and buzzing alarm 
and remote alarm 
function will be 
canceled after 
“Slience” is pressed. If 
the machine is still in 
alarm status after Fd 
time (user parameter 
can be set and the 
default value is 30min), 
the buzzing alarm and 
remote alarm will restart. 

When power outage or the 
power switch is not turned on, 
the outage alarm occurs. In case 
of power outage, the computer 
board will be powered by 
standby rechargeable battery. 

The 

power 

indicator 

lights of 

alarm 

light flash 

at the 
same 

time 

Display the average 
temperature of upper 
and lower sensors in 
refrigerator for 5s 
(after power outage, 
the average 
temperature of upper 
and lower sensor 
also display for 5s) 
and the screen turns 
black for 30s, which 
is alternative. 

Start remote alarm. 
Buzzing can be 
canceled after 
pressing “Slience”. If 
the refrigerator is still 
in alarm status after 
Fd time (user 
parameter can be set 
and the default value 
is 30min), the buzzing 
alarm will restart until 
the outage is solved. 

Upper sensor is in short circuit 
or broken circuit, fault code: E1 

Display board 
displays the 
alternation between 
the temperature of 
lower sensor and E1, 
wherein, E1 
displaying for 2s 
while the temperature 
displaying for 6s. 

Lower sensor is in short circuit 
or broken circuit, fault code: E2 

Start remote alarm. 
Pressing “Slience” will 
cancel the functions of 
buzzer and remote 
alarm. If it is still in an 
alarming state after a 
period of Fd (user 
parameters can be set, 
the default value is 
30 min), the buzzer 
and remote alarm will 
be restarted. 

1Hz 

Continuously 

Buzzing 

1Hz 

Continuously 

Buzzing 

Alarm 

light 

flashes  Display board 

displays the 
alternation between 
the temperature of 
upper sensor and E2, 
wherein, E2 
displaying for 2s while 
the temperature 
displaying for 6s.

HXC-1308/1308B

Temperature recorder 

HXC-158/158B/358/358B/608/608B

Refrigerator temperature recorders of 6-inch standard panel can be chosen for 
HXC-158/158B/358/358B/608/608B to record the temperature change in box.  
   Installation and Operation 
In order to operate and use the recorder in a right way, please follow the steps below: 
1. Open the door of recorder. You can see the recorder; 
2. Switch on the 9V DC-battery on the right corner of the recorder. The battery is emergency power 
supply; 
3. Examine the recording paper. If it needs to be changed, please change the recording paper 
according to the operation steps in P19; 
4. Pull off the plastic cap from recording pen and close the door of recorder 

The recorder will not operate before the temperature in the box reaches the 
recording scope. 

     

Power supply 

The recorder power is supplied by alternating current (AC) in normal condition when the refrigerator 
operates. If AC power supply breaks down, LED indicator lamp on the recorder flashes and alarm 
indicates power supply is abnormal. The recorder continues recording of temperature in box with 
power supply from emergency battery. Each emergency battery can sustain for about 30h 
(Note: Low-power battery shall be replaced in time, to avoid corrosion of battery button. In order to
save battery, please cut off battery button when the refrigerator is not used. In normal application
process, please install emergency battery to guarantee normal recording when the power is
outaged). LED indicator lamp of the recorder continues to flash until the main power (AC) is 
connected and the emergency battery is changed. When the power of emergency battery is low,
flashing of LED indicator lamp on the recorder indicates that the battery needs to be changed. 

Chart Adjustment Buttons

9 volt battery

Chart Adjustment Buttons

Imprinting Stylus

Date Reference Mark

Center Nut and Chain

Chart Paper

Summary of Contents for HXC-158B

Page 1: ...echnology Development Zone Qingdao 266510 P R China Web www haiermedical com Version 1st 2016 Dedicated code 0270501029 V13026 Read the Operation Manual carefully before using your appliance Keep the...

Page 2: ......

Page 3: ...ent from the figure in the manual due to product improvement The manual may be subject to changes without further notice Computer panel control digital display of temperature temperature display accur...

Page 4: ...ety Precautions Precautions for Use Product Installation Component Names Control Panel Application method Alarm Cleaning and Maintenance FAQ Refrigeration Schematic Circuit Diagram Specification 2 3 6...

Page 5: ...t to prevent personal injuries and refrigerator damage In case of flammable gas leakage such as coal gas shut off the valve of leaked gas open windows for ventilation and exhaust do not plug in or pul...

Page 6: ...gged into the socket to prevent fire Apply socket with ground wire to prevent electric shock If the socket fails to be grounded ground wires must be installed by technician Do not use the refrigerator...

Page 7: ...o children And the flammable foaming needs to be disposed by professional persons CP IP foaming materials are flammable need professional processing HXC 608B 6 Drawers HXC 608 Blood Bank Refrigerator...

Page 8: ...solvent such as gasoline Shut off the power as well as the battery switch if the refrigerator will be out of service for a long time 39 6 Precautions for Use Specification HXC 608 115V 60Hz Blood Ban...

Page 9: ...The refrigerator shall be grounded in a reliable manner If the socket is provided with grounding wire check if it is well grounded before operation If the socket is not provided with grounding wire a...

Page 10: ...ilure alarm 48 hours when battery is fully charged I Temperature controller USB function temperature recorder Optional optional Y Electric shock resistance type Power connect style Climate Type Foamin...

Page 11: ...control panel Internal fan Display panel Sensor Defrosting Control Lower temperature Upper temperature Alarm VCC GND Heating strip at the cabinet outlet Heating strip of center sill Door heating trans...

Page 12: ...the articles gradually into the refrigerator after complete examination of refrigerator 35 10 Circuit Diagram HXC 158 158B 358 358B 608 608B CN2 Cooling fan Compressor Defrosting heating strip CN2 CN...

Page 13: ...rature at one time If so the compressor may not stop for a long time with temperature reducing slowly which will reduce service life Articles must be put in batches and the temperature shall be reduce...

Page 14: ...ean data tested in the standard laboratory without noise in which the refrigerator without storage of articles is laid on the rubber pad and is steadily operated after closing the door to measure out...

Page 15: ...e for detailed information Purchasing of backup battery of the recorder Battery in the same specification and the same brand can be purchased in proper shops 9V DC In order to guarantee the reliabilit...

Page 16: ...pment and may lead to the occurrence of fault Compressor and other mechanical parts are in fully closed state and they do not need to be lubricated Clean the refrigerator Cleaning and Maintenance Comp...

Page 17: ...attery switch is on or off the remote alarm can be started The remote alarm port coordinates with the audible alarm on the refrigerator Thus the remote alarm can be stopped by pressing the key of buzz...

Page 18: ...or the refrigerator He is in charge of examining the running status of the refrigerator and making the record If the temperature inside the refrigerator is too high or too low or other abnormal phenom...

Page 19: ...ack to the normal display mode The original temperature set value 4 flickers Please do the next operation within 60s Otherwise the display panel will return to display TS SN Alarm items Buzzing Light...

Page 20: ...fter a period of Fd user parameters can be set the default value is 30 min the buzzer and remote alarm will be restarted 1Hz Continuously Buzzing 1Hz Continuously Buzzing Alarm light flashes Display b...

Page 21: ...til the pen returns to the paper At this moment the pen will pause at the most external temperature scaling line of the recording paper this scaling line does not have to be marked with temperature va...

Page 22: ...g the change of temperature in the box In order to use the disk recorder in a right and reasonable way please read the instruction bellow carefully Get familiar with your temperature recorder Adjustme...

Page 23: ...ction The computer panel has USB output function which can make use of U disk to export temperature and other test data Plug in U disk the machine begins to automatic identification and import data in...

Page 24: ...position can also be adjusted finely 3s later the scriber will slide to appointed scale according to the present perceived temperature then the calibration ends After retightening hold down bolt and...

Reviews: