5-14
Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
223
CENTRIFUGE PROBLEM
THE CENTRIFUGE COVER IS NOT LOCKED.
1. Close Centrifuge Cover.
2. Tighten cover knob completely.
3. Resume operation.
224
VALVE PROBLEM
THE PURPLE VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
225
VALVE PROBLEM
THE ORANGE VALVE WAS DETECTED OUT OF POSITION
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
226
VALVE PROBLEM
THE BLUE VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
227
VALVE PROBLEM
THE YELLOW VALVE WAS DETECTED OUT OF POSITION
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
228
VALVE PROBLEM
THE WHITE VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
229
VALVE PROBLEM
THE GREEN VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
248
thru
252
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
254
VALVE PROBLEM
THE RED VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
255
CONNECTOR
THE DPM FILTER HAS BEEN DISCONNECTED.
Please install the filter.
1. Attach filter.
2. Ensure that the filter is not wet.
3. Resume operation.
256
CONNECTOR
THE SPM FILTER HAS BEEN DISCONNECTED.
Please install the filter.
1. Attach filter.
2. Ensure that the filter is not wet.
3. Resume operation.
#