5-20
Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
407
NOTICE
THE PROTOCOL CARD DOES NOT MATCH DISPOSABLE.
The installed Protocol Card and disposable
do not match.
1. Check that the correct disposable
is installed.
2. Check that the correct Protocol Card
is inserted.
3. Clean the disposable ID window.
408
SELF TEST PROBLEM
ANTICOAGULANT AIR DETECTOR (ACAD) TEST FAILED.
Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
409
SELF TEST PROBLEM
BLOOD LINE AIR DETECTOR (BLAD) TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
410
SELF TEST PROBLEM
DONOR LINE AIR DETECTOR 1 (DLAD1) TEST FAILED.
Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
411
SELF TEST PROBLEM
DONOR LINE AIR DETECTOR 2 (DLAD2) TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
413
VALVE PROBLEM
PURPLE VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
414
VALVE PROBLEM
ORANGE VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
415
VALVE PROBLEM
RED VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
416
VALVE PROBLEM
BLUE VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
417
VALVE PROBLEM
YELLOW VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
418
VALVE PROBLEM
WHITE VALVE SELF TEST FAILED.
1. Toggle the valve manually.
2. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
#