6-8
Maintenance
P/N 39551-00, Rev. F
5. Close the slide clamp of the used waste bag. Remove the bag and discard it
according to the facility’s SOP for disposition of biohazardous wastes. Re-
place the bag with the spare and call Haemonetics to report the spill and re-
quest another waste bag.
Cleaning the
Optics Lens
Bowl Optics
The Bowl Optics should be cleaned monthly.
1. Wipe the Bowl Optics with a lint-free cloth moistened with water.
2. Dry the Bowl Optics with another lint-free cloth.
The window of the Bowl Optics located inside the Centrifuge may be cleaned
with a soft, damp, lint-free cloth (moistened with water only). The optic lens
should be cleaned monthly or when a spill occurs.
Warning!
In order to function properly, the optical sensor
lens must be clean and clear. A dirty or clouded lens could in-
terfere with proper operation of the sensor. Always clean the
optical lens after a blood spill.
Cleaning the
Line Sensor
Line Sensor
The line sensor should be cleaned at least once each month as well as any time
a spill has occurred.
1. Gently wash the groove of the line sensor with a lint-free cloth moistened
with clear, warm water.
2. Dry the groove with a lint-free cloth.
Warning!
The line sensor must be clean to function properly.
If it is not clean, it may interfere with the proper performance of
the apheresis procedure.
Cleaning the
Disposable
Identification
Window
Disposable Identification Window
The disposable identification window is located next to the blood and white
pump manifold. The disposable identification window should be cleaned
monthly.
1. Wipe the disposable identification window with a lint-free cloth moistened
with water.
2. Dry the disposable identification window with another lint-free cloth.
Summary of Contents for MCS+
Page 2: ...P N 39551 00 Rev F...
Page 10: ......
Page 11: ...P N 39551 00 Rev F Chapter One INTRODUCTION OVERVIEW OF THE MCS DEVICE 1 2...
Page 14: ......
Page 44: ......
Page 56: ......
Page 86: ......
Page 102: ......