2-14
System Description
P/N 39551-00, Rev. F
Pressing the UP/DOWN arrows raises or lowers the Blood Pump speed of the
current mode. Any change to the set speed made by using these keys is retained
throughout the procedure until another change is made.
Note:
The MCS+ device will automatically adjust the pump speeds in poor
flow conditions. A poor flow condition is indicated by the changing level of the
bar graph on the display screen and a change in the donor/patient flow indicator
lights. Wait until the pump speed is at a constant rate before readjusting pump
speed to donor/patient flow.
Cuff Key
Figure 2-8, Cuff Key
During all procedures, a Tourniquet Cuff placed above the venipuncture site will
automatically inflate during the Draw and deflate during Surge and Return. How-
ever, the CUFF key can be used during the Draw and Ready modes to release
tourniquet pressure. The current status of the Cuff is indicated by the light on the
CUFF key. When the light is on, the Cuff is inflated. (Cuff pressure is displayed
in Modify screen.)
Programming
Keys
Figure 2-9, Programming Keys
✍
Cuff
Modify
Yes
No
Save
Help
Haemo
Calculator
+
–