100
CENTRIFUGE PROBLEM
CENTRIFUGE SAFETY SWITCH NOT OPERATIONAL.
The centrifuge safety switch must be
operational to operate the MCS+ device.
1. Power off, and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
101
PROCESS PROBLEM
RETURN CYCLE TAKING TOO LONG.
Emptying bowl took longer than expected.
An occlusion may have occurred.
1. Check tubing at pumps and bowl.
2. Check bowl.
3. Resume operation.
102
PROCESS PROBLEM
BLAD DETECTED AIR EARLIER THAN EXPECTED.
BLAD DETECTED AIR EARLIER THAN EXPECTED
1. Check tubing installation in BLAD.
2. Open the RED valve and turn the BLOOD pump
CLOCKWISE to move air into the bowl.
3. Resume operation after air is cleared from tubing
back to bowl.
103
NO AC FLOW
LOW AC FLOW DETECTED BY DRIP MONITOR.
1. Check that the AC bag contains fluid.
2. Check AC tubing for occlusions.
3. Check AC tubing for clamps.
4. Check Drip Chamber installation.
5. Check Drip Chamber fluid level.
6. Resume operation.
104
NO AC FLOW
HIGH AC FLOW DETECTED BY DRIP MONITOR.
1. Check tubing at AC Pump.
2. Check Drip Chamber installation.
3. Check Drip Chamber fluid level.
4. Resume operation.
105
NO AC FLOW
NO AC FLOW DETECTED BY DRIP MONITOR.
1. Check that the AC bag contains fluid.
2. Check donor tubing for occlusions.
3. Check donor tubing for closed clamps.
4. Check AC tubing for occlusions.
5. Check AC tubing for closed clamps.
6. Resume operation.
106
SYSTEM PROBLEM
NO CUFF PRESSURE DETECTED.
1. Check that cuff is attached.
2. Check cuff for leaks.
3. Resume operation.
107
PROCESS PROBLEM
NO AIR / PLASMA INTERFACE DETECTED.
1. Check pump tubing installation.
2. Check tubing for occlusions.
3. Check venipuncture.
4. Press RETURN to continue.
175
ACAD DETECTS AIR
ANTICOAGULANT AIR DETECTOR
ANTICOAGULANT AIR DETECTOR
1. Check tubing installation in ACAD.
2. If air is present, turn AC pump CLOCKWISE to
move air past ACAD.
3. Call Haemonetics Technical Service if problem
persists.
3. Resume operation after air is cleared from ACAD.
#