55
Modbus Protocol
Setup/Analog Output
1/PID
PID Mode
40018
Unsigned
Integer
1
R/W
Sets the PID mode (0 = auto; 1 =
manual)
Setup/Analog Output
1/PID
PID Manual
Set
40019
Float
2
R/W
Sets the PID manual output value
(0.0 to 100.0%)
Setup/Analog Output
1/PID
PID Setpoint
40021
Float
2
R/W
Sets the PID setpoint
Setup/Analog Output
1/PID
PID Phase
40023
Unsigned
Integer
1
R/W
Sets the PID phase (0 = Direct;
1 = Reverse)
Setup/Analog Output
1/PID
PID
Proportional
Band
40024
Float
2
R/W
Sets the PID proportional band
Setup/Analog Output
1/PID
PID Integral
Time
40026
Unsigned
Integer
1
R/W
Sets the PID integral time (min)
Setup/Analog Output
1/PID
PID
Derivative
Time
40027
Unsigned
Integer
1
R/W
Sets the PID derivative time (min)
Setup/Analog Output 2
Source
40028
Unsigned
Integer
1
R/W
Selects data source for this output
(0 = None; 2 = Sensor; 4 = Calculation)
Setup/Analog Output 2
Sensor
Select
40029
Unsigned
Integer
1
R/W
Selects sensor source when Source =
Sensor (0 = Sensor1; 1 = Sensor2)
Setup/Analog Output 2
Measuremen
t Select
40030
Unsigned
Integer
1
R/W
Selects measurement on the Sensor
(0 = Meas1 .. 3 = Meas4)
Setup/Analog Output 2
Type
40031
Unsigned
Integer
1
R/W
Selects output type (0 = Linear output;
1 = PID control)
Setup/Analog Output 2
Transfer
Value
40032
Float
2
R/W
Sets the transfer value
Setup/Analog Output 2
Filter
40034
Unsigned
Integer
1
R/W
Sets the output filter value in seconds
(0 to 120 sec)
Setup/Analog Output 2
0mA - 4mA
Select
40035
Unsigned
Integer
1
R/W
Selects 0mA/4mA for min output
(0 = 0mA; 1 = 4mA)
Setup/Analog Output
2/Linear
Min Setting
40036
Float
2
R/W
Sets the min output value
Setup/Analog Output
2/Linear
Max Setting
40038
Float
2
R/W
Sets the max output value
Setup/Analog Output
2/PID
Mode
40040
Unsigned
Integer
1
R/W
Sets the PID mode (0 = auto; 1 =
manual)
Setup/Analog Output
2/PID
Manual Set
40041
Float
2
R/W
Sets the PID manual output value
(0.0 to 100.0%)
Setup/Analog Output
2/PID
Setpoint
40043
Float
2
R/W
Sets the PID setpoint
Setup/Analog Output
2/PID
Phase
40045
Unsigned
Integer
1
R/W
Sets the PID phase
(0 = Direct; 1 = Reverse)
Setup/Analog Output
2/PID
Proportional
Band
40046
Float
2
R/W
Sets the PID proportional band
Setup/Analog Output
2/PID
Integral Time
40048
Unsigned
Integer
1
R/W
Sets the PID integral time (min)
Setup/Analog Output
2/PID
Derivative
Time
40049
Unsigned
Integer
1
R/W
Sets the PID derivative time (min)
Table 11 sc100 Modbus Registers (continued)
Group Name
Tag Name
Register
#
Data
Type
Length
R/W
Description
Summary of Contents for sc100
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 8: ...Visit us at www hach com...
Page 28: ...Visit us at www hach com...
Page 44: ...Visit us at www hach com...
Page 50: ...Visit us at www hach com...
Page 62: ...Visit us at www hach com...
Page 64: ...Visit us at www hach com...