16
Installation
Figure 10
Removing Voltage Barrier
3.3.1 Installation in Conduit
In hard-wired electrical applications, the power and safety ground service drops for the
instrument must be 18 to 12 AWG. See
Figure 11 on page 16
for strain relief and conduit
opening sealing plug information. See
section 3.3.3 on page 17
for wiring information.
3.3.2 Installation Using a Power Cord
DANGER
Use of a power cord is not acceptable in Class 1, Division 2 Hazardous Location
Installation (see
Hazardous Location Installation Control Drawing on page 10
).
A sealing-type strain relief to maintain the NEMA 4X/IP66 environmental rating and a
power cord less than 3 meters (10 feet) in length with three 18-gauge conductors
(including a safety ground wire) can be used, see
Replacement Parts and Accessories on
page 43
. See
Figure 11 on page 16
for strain relief and conduit opening sealing plug
assembly. See
section 3.3.3 on page 17
for wiring information.
Figure 11
Using the Optional Strain Relief and Conduit Plug
1.
High voltage barrier
2.
Unsnap the barrier latch then pull out to remove the barrier.
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY B
RELAY A
RELAY A
1
1
+ DATA
+ DATA
+ OUT 2
+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V
+ V
+ OUT 1
+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY B
RELAY A
RELAY A
1
1
+ DATA
+ DATA
+ OUT 2
+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V
+ V
+ OUT 1
+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
2
1
1.
Power cord strain relief
2.
Conduit strain relief
3.
Conduit opening sealing plug
2
3
Summary of Contents for sc100
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 8: ...Visit us at www hach com...
Page 28: ...Visit us at www hach com...
Page 44: ...Visit us at www hach com...
Page 50: ...Visit us at www hach com...
Page 62: ...Visit us at www hach com...
Page 64: ...Visit us at www hach com...