Modbus Protocol
Setup/Relay 3/Alarm
Low
Deadband
40141
Float
2
R/W
Sets the low alarm deadband
Setup/Relay 3/Alarm
On Delay
40143
Unsigned
Integer
1
R/W
Sets the on delay time
Setup/Relay 3/Alarm
Off Delay
40144
Unsigned
Integer
1
R/W
Sets the off delay time
Setup/Relay 3/Control
Setpoint
40145
Float
2
R/W
Sets the controller setpoint
Setup/Relay 3/Control
Phase
40147
Unsigned
Integer
1
R/W
Sets the controller phase
(0 = Low; 1 = High)
Setup/Relay 3/Control
Deadband
40148
Float
2
R/W
Sets the controller deadband
Setup/Relay 3/Control
Overfeed
Timer
40150
Unsigned
Integer
1
R/W
Sets the overfeed timer value (mins)
Setup/Relay 3/Control
On Delay
40151
Unsigned
Integer
1
R/W
Sets the on delay time (sec)
Setup/Relay 3/Control
Off Delay
40152
Unsigned
Integer
1
R/W
Sets the off delay time (sec)
Setup/Relay 3/Control
Reset
Overfeed
Timer
40153
Unsigned
Integer
1
R/W
Resets the overfeed timer
Setup/Relay 3/Event
Setpoint
40154
Float
2
R/W
Sets the event setpoint
Setup/Relay 3/Event
Phase
40156
Unsigned
Integer
1
R/W
Sets the event phase (0 = Low; 1 =
High)
Setup/Relay 3/Event
Deadband
40157
Float
2
R/W
Sets the event deadband
Setup/Relay 3/Event
On Max Time
40159
Unsigned
Integer
1
R/W
Sets the max on time (mins)
Setup/Relay 3/Event
On Min Time
40160
Unsigned
Integer
1
R/W
Sets the min on time (mins)
Setup/Relay 3/Event
Off Max Time
40161
Unsigned
Integer
1
R/W
Sets the max off time (mins)
Setup/Relay 3/Event
Off Min Time
40162
Unsigned
Integer
1
R/W
Sets the min off time (mins)
Setup/Relay 3/Timer
Hold Type
40163
Unsigned
Integer
1
R/W
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Setup/Relay 3/Timer
Sensor
Select
40164
Unsigned
Integer
1
R/W
Selects which Sensor outputs are being
held/transferred during the timers on
time (this is used when Hold type is set
for single Sensor)
Setup/Relay 3/Timer
Hold Mode
40165
Unsigned
Integer
1
R/W
Selects hold outputs vs. set transfer
value during timers on time
Setup/Relay 3/Timer
Duration
Time
40166
Unsigned
Integer
1
R/W
Sets the timer on duration time (sec)
Setup/Relay 3/Timer
Period Time
40167
Unsigned
Integer
1
R/W
Sets the period between timer on events
(mins)
Setup/Relay 3/Timer
Off Delay
40168
Unsigned
Integer
1
R/W
Sets the time the affected Sensor
outputs are held/transferred after the
timer turns off (sec)
Setup/Relay 3/Status
Level
40169
Unsigned
Integer
1
R/W
Sets the status level which will trigger
the relay
Table 11 sc100 Modbus Registers (continued)
Group Name
Tag Name
Register
#
Data
Type
Length
R/W
Description
Summary of Contents for sc100
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 8: ...Visit us at www hach com...
Page 28: ...Visit us at www hach com...
Page 44: ...Visit us at www hach com...
Page 50: ...Visit us at www hach com...
Page 62: ...Visit us at www hach com...
Page 64: ...Visit us at www hach com...