18
Installation
The DC power source that supplies power to the 24 VDC powered sc100 must maintain
voltage regulation within the specified 24 VDC –15% +20% voltage limits. The DC power
source must also provide adequate protection against surges and line transients.
Figure 13
sc100 Wiring Connections
Table 2 AC Power Wiring Information (sc100 AC powered model only)
Terminal Number
Terminal Description
Wire Color Code for North America
Wire Color Code for Europe
1
Hot (L1)
Black
Brown
2
Neutral (N)
White
Blue
3
Protective Earth (PE)
Green
Green w/yellow tracer
Table 3 DC Power Wiring Information (sc100 24 VDC powered model only)
Terminal Number
Terminal Description
Wire Color Code for North America
Wire Color Code for Europe
1
+24 V dc
Red
Red
2
24 V dc return
Black
Black
3
Protective Earth (PE)
Green
Green w/yellow tracer
1.
J1—Network connector
8.
Sensor connector
2.
J2—Header for optional network interface card
9.
Sensor connector
3.
J5—Relay A connector
10. J6—Analog output (4–20 mA) connector
4.
J6—Relay B connector
11. J5—Sensor connector for hard-wiring
5.
J7—Relay C connector
12. Position for network interface card
6.
Fuses (F1, F2)
13. Service port
7.
J8—Power connections
14. Sensor terminator selector/service port configuration
a.
AC Power connection (AC powered sc100 model only)
b.
DC Power connection (24 VDC sc100 model only)
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
6
7
7a
7b
14
12
13
3
1
2
8
9
4
5
11
10
+24
VDC
+24
VDC
–24
VDC
–24
VDC
Summary of Contents for sc100
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 8: ...Visit us at www hach com...
Page 28: ...Visit us at www hach com...
Page 44: ...Visit us at www hach com...
Page 50: ...Visit us at www hach com...
Page 62: ...Visit us at www hach com...
Page 64: ...Visit us at www hach com...