21
Installation
Figure 16
Alarm and Relay Connections
3.4.2
Connecting the Analog Outputs
DANGER
Explosion hazard. Do not connect or disconnect electrical components or circuits
to the equipment unless power has been switched off or the area is known to be
non-hazardous.
DANGER
For Class 1, Division 2 Hazardous Location installations, refer to the Control
Drawing (
Figure 2 on page 10
) for permanent connection requirements for the
analog output.
Two isolated analog outputs (1 and 2) are provided, see
Figure 17
. Each output can be set
to 0–20 or 4–20 mA, and can be assigned to represent the measured parameter or
secondary measurement such as temperature. Make connections with twisted-pair
shielded wire and connect the shield at the controlled component end or at the control
loop end. Do not connect the shield at both ends of the cable. Use of non-shielded
cable may result in radio frequency emission or susceptibility levels higher than allowed.
Maximum loop resistance is 500 ohm. Refer to the sensor manual for output software
setup.
Make wiring connections at the analyzer end as shown in
Table 4
and
Figure 17
.
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
J5
J5
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
J5
J5
J5
J6
J7
3
2
1
RELAY A
RELAY B
RELAY C
Disconnect
Power
NC
COM
NO
Summary of Contents for sc100
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 8: ...Visit us at www hach com...
Page 28: ...Visit us at www hach com...
Page 44: ...Visit us at www hach com...
Page 50: ...Visit us at www hach com...
Page 62: ...Visit us at www hach com...
Page 64: ...Visit us at www hach com...