Normal operation of this device may require the use of chemicals or samples that are biologically
unsafe.
• Observe all cautionary information printed on the original solution containers and safety data
sheets prior to their use.
• Dispose of all consumed solutions in accordance with the local and national regulations and laws.
• Select the type of protective equipment suitable to the concentration and quantity of the dangerous
material being used.
If the wiper interval is correctly set for the application and the wiper profile is replaced regularly, more
cleaning of the measuring path is not necessary.
1.
Go to the main menu, then select
Sensor setup
.
2.
Select the applicable sensor, if necessary.
3.
Select
Maintenance
.
4.
Set the outputs on
Hold
.
5.
Remove the probe from the tank or flow through unit.
6.
Go back to the sensor menu, then select
Diagnostics/Test
>
Signals
.
7.
Rinse the probe with clean water.
As an alternative, use a window cleaner, grease remover or 5 % hydrochloric acid to clean the
measuring path based on the quantity and type of contamination. Soak the probe in the cleaning
solution for 5 to 10 minutes.
8.
Carefully clean the measuring path with distilled water.
Clean the measuring path until the
DExt1
and
DExt2
values are less than 15.
9.
Select
Back
to go back to
Diagnostics/Test
.
10.
Select
Back
again.
11.
Select
Return sensor to process
.
The probe goes back to measurement mode after the automatic wipe cleaning.
12.
Install the probe in the tank or flow through unit.
6.3 Replace the wiper blade
Replace the wiper blade when one or more of the conditions that follow occurs:
• After 25,000 cleaning cycles (A-B-A or B-A-B)
• After 50,000 cleaning single movements (A-B or B-A)
• if the wiper blade has damage or does not operate correctly.
1.
Remove the probe from the tank or flow through unit.
2.
Go to the maintenance menu as follows:
• SC200 and SC1000 Controllers—Go to the main menu, then select
Sensor setup
>
Maintenance
.
• SC4500 Controller and Claros interface—Push the tile of the device and select
Device menu
>
Maintenance
.
3.
Select
Wiper replacement
.
4.
Complete the instructions on the display. When asked, replace the wiper blade. Refer to the
illustrated steps in
and
.
English
19
Summary of Contents for NT3100sc
Page 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Page 11: ...Figure 3 Bracket installation English 11...
Page 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Page 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Page 104: ...2 1 2 1 1 2 1 2 104...
Page 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Page 109: ...3 109...
Page 110: ...4 5 110...
Page 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Page 126: ...2 1 1 2 1 2 126...
Page 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Page 131: ...3 2 3 5 4 3 131...
Page 132: ...4 5 132...
Page 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Page 140: ...8 1 mm 2 mm 9 5 mm 140...
Page 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Page 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Page 149: ...2 1 1 2 1 2 UV 149...
Page 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Page 154: ...4 3 154...
Page 155: ...4 5 155...
Page 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Page 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Page 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Page 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Page 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Page 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Page 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Page 176: ...3 3 1 OFF 3 2 3 5 4 176...
Page 177: ...3 177...
Page 178: ...4 5 178...
Page 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Page 193: ......