Table 1 General specifications (continued)
Specification
Details
Fittings
Sample line and sample bypass drain: 6-mm OD push-to-connect fitting for plastic
tubing
Chemical and case drains: 7/16-in. ID slip-on fitting for soft plastic tubing
Certifications
CE-compliant, CB, cETLus, TR CU-compliant, RCM, KC
Table 2 Sample requirements
Specification
Details
Sample pressure
0.2 to 6 bar (3 to 87 psi)
Sample flow rate
100 to 150 mL/minute (6 to 9 L/hour)
Sample temperature
5 to 45 °C (41 to 113 °F)
Sample pH
Analyzers without cationic pump: 6 to 10 pH
Analyzers with cationic pump: 2 to 10 pH
Sample acidity (equivalent CaCO
3
)
Analyzers without cationic pump: Less than 50 ppm
Analyzers with cationic pump: Less than 250 ppm
Suspended solids in sample
Less than 2 NTU, no oil, no grease
Table 3 Measurement specifications
Specification
Details
Electrode type
Sodium ISE (ion specific electrode) electrode and reference electrode with KCl
electrolyte
Measurement range
Analyzers without cationic pump: 0.01 to 10,000 ppb
Analyzers with cationic pump: 0.01 ppb to 200 ppm
Accuracy
Analyzers without cationic pump:
• 0.01 ppb to 2 ppb: ± 0.1 ppb
• 2 ppb to 10,000 ppb: ± 5%
Analyzers with cationic pump:
• 0.01 ppb to 40 ppb: ± 2 ppb
• 40 ppb to 200 ppm: ± 5%
Precision/Repeatability
Less than 0.02 ppb or 1.5% (the larger value) with ± 10 °C (50 °F) sample
difference
Interference phosphate 10 ppm Measurement interference is less than 0.1 ppb
Response time
Refer to
Stabilization time
Startup: 2 hours; Sample temperature variation: 10 minutes from 15 to 30 °C
(59 to 86 °F)
Use the optional heat exchanger when the temperature difference between
samples is more than 15 °C (27 °F).
Calibration time
50 minutes (typical)
Calibration
Automatic calibration: known addition method; Manual calibration: 1 or 2 points
Minimum detection limit
0.01 ppb
4
English
Summary of Contents for NA5600 sc Na+
Page 17: ...English 17 ...
Page 34: ...Figure 20 DIPA bottle installation Analyzer without enclosure 34 English ...
Page 54: ...54 Deutsch ...
Page 62: ...Abbildung 10 Führen des Kabels durch einen internen Anschlussstecker 62 Deutsch ...
Page 72: ...Abbildung 20 Installation der DIPA Flasche Analysator ohne Gehäuse 72 Deutsch ...
Page 92: ...92 Italiano ...
Page 100: ...Figura 10 Far passare il cavo attraverso il tappo di una porta interna 100 Italiano ...
Page 110: ...Figura 20 Installazione del flacone di DIPA Analizzatore senza alloggiamento 110 Italiano ...
Page 130: ...130 Français ...
Page 138: ...Figure 10 Acheminement du câble par un connecteur de port interne 138 Français ...
Page 148: ...Figure 20 Installation du flacon de DIPA Analyseur sans boîtier 148 Français ...
Page 167: ...Español 167 ...
Page 175: ...Figura 10 Pasar el cable a través de un tapón de puerto interno Español 175 ...
Page 185: ...Figura 20 Instalación de una botella de DIPA en un analizador sin carcasa Español 185 ...
Page 202: ...3 5 1 Remover a tampa de acesso eléctrico Consulte os passos ilustrados abaixo 202 Português ...
Page 204: ...204 Português ...
Page 212: ...Figura 10 Passar o cabo através de uma tampa de porta interna 212 Português ...
Page 222: ...Figura 20 Instalação do frasco de DIPA Analisador sem estrutura 222 Português ...
Page 241: ...Nederlands 241 ...
Page 249: ...Afbeelding 10 Steek de kabel door de tule van een interne poort Nederlands 249 ...
Page 259: ...Afbeelding 20 DIPA fles plaatsen Analyser zonder behuizing Nederlands 259 ...
Page 279: ...Polski 279 ...
Page 287: ...Rysunek 10 Poprowadzenie przewodu przez zatyczkę portu wewnętrznego Polski 287 ...
Page 297: ...Rysunek 20 Instalacja butelki z DIPA analizator bez obudowy Polski 297 ...
Page 314: ...3 5 1 Ta bort el åtkomstskyddet Se de illustrerade stegen som följer 314 Svenska ...
Page 316: ...316 Svenska ...
Page 333: ...Figur 20 Installation av DIPA flaska analysator utan kapsling Svenska 333 ...
Page 351: ...Suomi 351 ...
Page 368: ...Kuva 20 DIPA pullon asennus analysaattori jossa ei ole koteloa 368 Suomi ...
Page 387: ...Magyar 387 ...
Page 395: ...10 ábra Vezesse át a kábelt az egyik belső csatlakozó dugóján Magyar 395 ...
Page 405: ...20 ábra DIPA palack beszerelése ház nélküli analizátor Magyar 405 ...
Page 425: ...Русский 425 ...
Page 433: ...Рисунок 10 Пропустите кабель через заглушку внутреннего порта Русский 433 ...
Page 443: ...Рисунок 20 Установка бутылки DIPA Анализатор без корпуса Русский 443 ...
Page 461: ...Türkçe 461 ...
Page 478: ...Şekil 20 DIPA şişesinin takılması Muhafazasız analiz cihazı 478 Türkçe ...
Page 482: ......
Page 483: ......