Figure 1 Product overview
1
Power input and communication
connector, 10-pin
1
7
Relative humidity (RH) and
temperature sensor port
2
Vacuum source fitting (or quick-
connect fitting)
8
Vacuum source fitting (alternate
location)
3
Sample air inlet fitting
9
Connection indicator lights
2
(
4
Status indicator light (
10
Ethernet RJ45 connector
2
5
Service port and optional light stack
port
11
Power input connector, 5-pin
2
6
DIP switch, network address
1
1
All units except Ethernet
2
Ethernet units only
Table 1 Status indicator light
Color
Indication
System status
Green
Flashing (3 seconds)
Normal, sampling
On
Normal, not sampling
Table 1 Status indicator light (continued)
Color
Indication
System status
Blue
On
Sensor failure
One short flash, one long flash
Air flow failure
Flashing
Communication failure
Red
On or flashing
Count alarm
Yellow
On
Initialization
Flashing
Count alert
1
Purple
Flashing
Setup utility is being used
1
The user-supplied central monitoring software can be used to make the yellow
light flash when a count alert occurs with ModBus protocol, not FX protocol.
The count alert settings are selected with the central monitoring software.
Table 2 Ethernet indicator lights
Color
Indication
Status
Yellow
On
Connected
Green
Off
10Base-T
On
100Base-T
Isokinetic probe
For the best accuracy in laminar flow environments, always use the
supplied isokinetic probe with this instrument. The velocity of air in the
probe is similar to that of a typical vertical or horizontal laminar flow
environment such as a clean room or clean hood. The supplied
isokinetic probe supplies the same vertical (or horizontal) flow speed of
the air in order to collect representative samples of the clean room
laminar flow for the instrument. Refer to
for a comparison of
sampling with and without the isokinetic probe.
6
English
Summary of Contents for MET ONE 6003
Page 2: ...English 3 Fran ais 24 Espa ol 48 72 91 113 2...
Page 13: ...Figure 9 5 pin connector wiring Figure 10 10 pin connector wiring English 13...
Page 75: ...1 3 1 1 ModBus FX 2 Status 10Base T 100Base T 2 2 1 3 2 4 3 3 75...
Page 77: ...3 m 10 ft 3 m 10 ft 1 m 3 m 10 ft schedule 80 PVC Cobolite DIN DIN 5 77...
Page 78: ...5 DIN 79 6 90 90 T 78...
Page 81: ...9 5 10 10 81...
Page 83: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 83...
Page 89: ...3 PC CSV 15 1 1 7 cfm L min 2 8 cfm L min 3 9 RH 4 10 5 11 6 1 89...
Page 90: ...9 9 1 90 X 90 X FMS 90...
Page 95: ...2 2 1 3 2 4 3 3 3 1 3 5 2 4 3 0 0 1 cfm 0 3 m 0 5 m F 1 1 0 cfm 0 5 m N 95...
Page 97: ...3 m 3 m 1 m 3 m 80 PVC Cobolite DIN DIN 5 97...
Page 98: ...5 DIN 99 6 90 90 T 98...
Page 99: ...6 1 3 2 90 4 T 2 3 m2 1 2 2 95 30 cm 30 cm 4 4 Hytrel Bevaline Tygon Hytrel Bevaline Tygon 99...
Page 112: ...HACH Japan 9 9 1 112 X 112 X FMS 1 112...
Page 114: ...5 1 CE 1 ISO21501 4 CD 114...
Page 119: ...DIN DIN 5 5 DIN 119...
Page 120: ...120 6 90 90 T 6 1 3 2 90 4 T 2 3 m2 25 ft2 1 120...
Page 123: ...9 5 10 10 123...
Page 125: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 125...
Page 133: ...9 9 1 133 X 133 X FMS 1 133...
Page 134: ......
Page 135: ......