13.
On the Basic Setup tab, change the settings back to the previous
values.
14.
Click
Save Settings
.
Table 7 Output voltage with 100, 250 and 500-ohm resistors
Sample count
100 Ω
250 Ω
500 Ω
0
0.40 V
1.00 V
2.00 V
100
0.56 V
1.40 V
2.80 V
200
0.72 V
1.80 V
3.60 V
300
0.88 V
2.20 V
4.40 V
400
1.04 V
2.60 V
5.20 V
500
1.20 V
3.00 V
6.00 V
600
1.36 V
3.40 V
6.80 V
700
1.52 V
3.80 V
7.60 V
800
1.68 V
4.20 V
8.40 V
900
1.84 V
4.60 V
9.20 V
1000
2.00 V
5.00 V
10.00 V
RS485 serial output with Modbus RTU protocol
Instruments with the RS485 Modbus communication option use industry-
standard Modbus RTU protocol. In this communication mode, a series of
registers hold data about measurement results and operation
parameters.
When a ModbusTCP connection is made, the user can use all the
configuration options in the Modbus register map. Refer to the company
website for the Modbus register map. Write drivers to communicate with
the instrument through these registers with the Modbus RTU protocol.
The RS485 serial network circuit supplies communications for a
maximum of 32 instruments and a control computer. Only one
instrument can transmit data at a time. Each instrument must have a
unique instrument address.
1.
Turn the instrument over. The DIP switch is on the bottom of the
instrument.
2.
Change the DIP switch setting to select a unique network address for
the instrument. Refer to
Note: Address 0 can only be used with FXB protocol. Address 0 is reserved for
use as a broadcast address for Modbus RTU. If address 0 is set with Modbus
protocol, the instrument will use address 1.
Table 8 DIP switch settings for network address
Network address
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
0
Off
Off
Off
Off
Off
1
On
Off
Off
Off
Off
2
Off
On
Off
Off
Off
3
On
On
Off
Off
Off
4
Off
Off
On
Off
Off
5
On
Off
On
Off
Off
6
Off
On
On
Off
Off
7
On
On
On
Off
Off
8
Off
Off
Off
On
Off
9
On
Off
Off
On
Off
10
Off
On
Off
On
Off
11
On
On
Off
On
Off
12
Off
Off
On
On
Off
13
On
Off
On
On
Off
14
Off
On
On
On
Off
15
On
On
On
On
Off
16
Off
Off
Off
Off
On
17
On
Off
Off
Off
On
20
English
Summary of Contents for MET ONE 6003
Page 2: ...English 3 Fran ais 24 Espa ol 48 72 91 113 2...
Page 13: ...Figure 9 5 pin connector wiring Figure 10 10 pin connector wiring English 13...
Page 75: ...1 3 1 1 ModBus FX 2 Status 10Base T 100Base T 2 2 1 3 2 4 3 3 75...
Page 77: ...3 m 10 ft 3 m 10 ft 1 m 3 m 10 ft schedule 80 PVC Cobolite DIN DIN 5 77...
Page 78: ...5 DIN 79 6 90 90 T 78...
Page 81: ...9 5 10 10 81...
Page 83: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 83...
Page 89: ...3 PC CSV 15 1 1 7 cfm L min 2 8 cfm L min 3 9 RH 4 10 5 11 6 1 89...
Page 90: ...9 9 1 90 X 90 X FMS 90...
Page 95: ...2 2 1 3 2 4 3 3 3 1 3 5 2 4 3 0 0 1 cfm 0 3 m 0 5 m F 1 1 0 cfm 0 5 m N 95...
Page 97: ...3 m 3 m 1 m 3 m 80 PVC Cobolite DIN DIN 5 97...
Page 98: ...5 DIN 99 6 90 90 T 98...
Page 99: ...6 1 3 2 90 4 T 2 3 m2 1 2 2 95 30 cm 30 cm 4 4 Hytrel Bevaline Tygon Hytrel Bevaline Tygon 99...
Page 112: ...HACH Japan 9 9 1 112 X 112 X FMS 1 112...
Page 114: ...5 1 CE 1 ISO21501 4 CD 114...
Page 119: ...DIN DIN 5 5 DIN 119...
Page 120: ...120 6 90 90 T 6 1 3 2 90 4 T 2 3 m2 25 ft2 1 120...
Page 123: ...9 5 10 10 123...
Page 125: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 125...
Page 133: ...9 9 1 133 X 133 X FMS 1 133...
Page 134: ......
Page 135: ......