Table 6 Warning list (continued)
Warning
Description
Resolution
The device is not
calibrated.
The sensor has not been calibrated.
Calibrate the sensor.
Replace a sensor.
The Sensor days counter is more
than the interval selected for sensor
replacement. Refer to
on page 8.
Replace the sensor. Reset the Sensor days
counter on the Diagnostics/Test > Reset
menu (or Diagnostics/Test > Counter menu.
Calibration is in
progress…
A calibration was started but not
completed.
Return to calibration.
Outputs on hold
During calibration, the outputs were
set to hold for a selected time.
The outputs will become active after the
selected time period. As an alternative,
remove and then supply power to the
controller.
Linear temperature
compensation is out of
range.
The user-defined linear temperature
compensation is out of range.
The value must be between 0 and 4%/°C; 0 to
200 °C.
Temperature
compensation table is
out of range.
The user-defined temperature
compensation table is out of range.
The temperature is above or below the
temperature range defined by the table.
6.6 Event list
The diagnostic bar shows current activities such as configuration changes, alarms, warning
conditions, etc. A list of possible events is shown in
. Previous events are recorded in the
event log, which can be downloaded from the controller. Refer to the controller documentation for
data retrieval options.
Table 7 Event list
Event
Description
Calibration ready
The sensor is ready for calibration.
The calibration is OK.
The current calibration is good.
The time has expired.
The stabilization time during calibration expired.
The calibration failed.
The calibration failed.
The calibration is high.
The calibration value is above the upper limit.
K is out of range.
The cell constant K is out of range for the current calibration.
The reading is unstable.
The reading during calibration was unstable.
Change in configuration
float value
The configuration was changed—floating point type.
Change in configuration
text value
The configuration was changed—text type.
Change in configurationint value
The configuration was changed—integer value type.
Change in configuration
The configuration was reset to the default options.
Power is on.
The power was turned on.
ADC failure
The analog to digital conversion failed (hardware failure).
Flash erase
The flash memory was erased.
Temperature
The recorded temperature is too high or too low (-20 to 200 °C).
The sample calibration was started.
Start of calibration for conductivity
English
17
Summary of Contents for D3433B8.99
Page 218: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 218...
Page 219: ...1 1 0 5 4 CIP 2 0 75 5 3 Teflon 0 75 6 3 3 1 1 2 232 224 2 T 219...
Page 220: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 220...
Page 224: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 14 15 13 ID 224...
Page 225: ...4 4 3 1 2 3 4 5 6 OK 7 OK 8 9 OK 10 4 4 4 1 2 3 1 4 225...
Page 226: ...5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 4 4 6 1 2 OK 226...
Page 227: ...4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus 5 1 2 7 25 psi 50 kPa MSDS SDS 5 1 Borax 227...
Page 229: ...6 3 4 4 sc sc sc sc sc 6 4 229...
Page 231: ...6 100 C 0 C 221 0 4 C 0 200 C 6 6 7 7 K K 231...
Page 290: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 290...
Page 291: ...1 1 0 5 4 CIP 2 0 75 5 3 0 75 6 3 3 1 1 2 304 296 2 291...
Page 292: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 4 1 292...
Page 296: ...4 4 50 MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 15 14 13 296...
Page 297: ...4 4 3 1 2 3 4 0 5 6 7 8 9 10 4 4 4 1 2 3 1 4 297...
Page 298: ...5 6 7 8 9 10 4 4 5 1 2 3 4 5 1 6 7 4 4 6 1 2 298...
Page 299: ...4 4 7 1 2 3 4 4 5 Modbus Modbus 5 1 2 50 MSDS SDS 5 1 299...
Page 300: ...1 2 2 3 3 4 5 5 5 2 3 6 6 6 1 6 2 299 1 2 3 3 1089 1106 23 27 C 5 5 300...
Page 301: ...6 3 4 4 sc sc sc sc sc 6 4 301...
Page 303: ...6 100 C 0 C 293 0 4 C 0 200 C 6 6 7 7 K K 303...
Page 376: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 376...
Page 377: ...1 1 0 5 in 4 CIP 2 0 75 in 5 3 Teflon 0 75 in 6 3 3 1 1 2 390 382 Mounting examples 2 377...
Page 378: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 378...
Page 382: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T 4 4 2 sc 1 2 382...
Page 383: ...3 4 15 60 14 14 15 13 ID 4 4 3 1 2 3 4 0 5 6 OK 7 OK 8 9 OK OK 10 4 4 4 383...
Page 384: ...1 2 3 1 4 5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 384...
Page 385: ...4 4 6 1 2 OK 4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus Web 5 1 2 7 25 psi 50 kPa MSDS SDS 385...
Page 387: ...3 6 3 4 4 sc sc service sc sc sc 387...
Page 389: ...6 5 6 6 300 000 300 000 100 C 0 C 379 0 4 C 0 200 C 6 6 7 389...
Page 390: ...7 Event OK K K ADC flash Flash 20 200 C 7 390...
Page 430: ...2 1 2 2 2 SC4500 2 3 1 1 1 0 5 4 2 0 75 5 3 0 75 6 430...
Page 431: ...3 3 1 1 2 442 435 2 T 2 1 T T NPT 4 7 2 T NPT 5 3 6 431...
Page 438: ...5 1 2 7 25 50 MSDS SDS 5 1 1 2 3 4 5 5 5 6 6 6 1 438...
Page 444: ......
Page 445: ......