4.2 Configure the sensor
Use the Settings menu to enter a sensor name and select the type of sensor. Change options for
measurement, calibration, data handling and storage.
1.
Select the main menu icon, then select
Devices
. A list of all of the available devices shows.
2.
Select the sensor and select
Device menu
>
Settings
.
3.
Select an option.
• For sensors connected to a conductivity module, refer to
.
• For sensors connected to an sc digital gateway, refer to
.
Table 1 Sensors connected to conductivity module
Option
Description
Name
Changes the name that corresponds to the sensor on the top of the measurement
screen. The name is limited to 16 characters in any combination of letters, numbers,
spaces or punctuation.
Sensor S/N
Lets the user enter the serial number of the sensor. The serial number is limited to
16 characters in any combination of letters, numbers, spaces or punctuation.
Measurement type
Changes the measured parameter to Conductivity (default), TDS (total dissolved solids),
Salinity or Resistivity. When the parameter is changed, all other configured settings are
reset to the default values.
Format
Changes the number of decimal places that are shown on the measurement screen to
Auto (default), X.XXX, XX.XX, XXX.X or XXXX. When Auto is selected, the decimal
places automatically change.
Unit
Changes the units for the selected measurement—conductivity: μS/cm (default), mS/cm,
μS/m, mS/m or S/m; resistivity: MΩ (default), kΩ, Ωcm or Ωm; TDS: ppm (default) or ppb;
salinity:
0
/
00
(ppt)
Temperature
Sets the temperature units to °C (default) or °F.
T-compensation
Adds a temperature-dependent correction to the measured value—Linear (default:
2.0%/°C, 25 °C), Ammonia, Temperature compensation table, None, Natural water or
Pure water.
When Temperature compensation table is selected, the user can enter x,y (%/°C, °C)
points in ascending order.
Note: The Natural water and Pure water options are not available when the Measurement type setting is
set to TDS.
TDS (total dissolved
solids)
Note: The TDS (total dissolved solids) setting is only available when the Measurement type setting is
set to TDS.
Sets the factor that is used to convert conductivity to TDS—NaCl (default) or Custom
(enter a factor between 0.01 and 99.99 ppm/µS, default: 0.49 ppm/µS).
Cell constant
parameters
Select the cell constant K.—Sets the cell constant range to 0.05, 0.5, 1.0 (default), 5.00,
10.0, or 0.01, 0.1, 1.0 Polymetron.
Cell constant—Enters the actual certified K value from the label on the sensor cable.
When the certified K value is entered, the calibration curve is defined.
Cable length
Sets the actual length of the sensor cable to improve measurement accuracy (default:
20 ft (Polymetron sensors default: 5 ft)).
Temperature
element
Sets the temperature element for automatic temperature compensation to PT100,
PT1000 (default) or Manual. If no element is used, set to Manual and set a value for
temperature compensation (default: 25 °C).
When Temperature element is set to PT100 or PT1000, refer to
on page 10 to set the Temperature factor setting.
Note: If Temperature element is set to Manual and the sensor is replaced or the sensor days are reset,
Temperature element automatically changes back to the default setting (PT1000).
8
English
Summary of Contents for D3433B8.99
Page 218: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 218...
Page 219: ...1 1 0 5 4 CIP 2 0 75 5 3 Teflon 0 75 6 3 3 1 1 2 232 224 2 T 219...
Page 220: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 220...
Page 224: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 14 15 13 ID 224...
Page 225: ...4 4 3 1 2 3 4 5 6 OK 7 OK 8 9 OK 10 4 4 4 1 2 3 1 4 225...
Page 226: ...5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 4 4 6 1 2 OK 226...
Page 227: ...4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus 5 1 2 7 25 psi 50 kPa MSDS SDS 5 1 Borax 227...
Page 229: ...6 3 4 4 sc sc sc sc sc 6 4 229...
Page 231: ...6 100 C 0 C 221 0 4 C 0 200 C 6 6 7 7 K K 231...
Page 290: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 290...
Page 291: ...1 1 0 5 4 CIP 2 0 75 5 3 0 75 6 3 3 1 1 2 304 296 2 291...
Page 292: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 4 1 292...
Page 296: ...4 4 50 MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 15 14 13 296...
Page 297: ...4 4 3 1 2 3 4 0 5 6 7 8 9 10 4 4 4 1 2 3 1 4 297...
Page 298: ...5 6 7 8 9 10 4 4 5 1 2 3 4 5 1 6 7 4 4 6 1 2 298...
Page 299: ...4 4 7 1 2 3 4 4 5 Modbus Modbus 5 1 2 50 MSDS SDS 5 1 299...
Page 300: ...1 2 2 3 3 4 5 5 5 2 3 6 6 6 1 6 2 299 1 2 3 3 1089 1106 23 27 C 5 5 300...
Page 301: ...6 3 4 4 sc sc sc sc sc 6 4 301...
Page 303: ...6 100 C 0 C 293 0 4 C 0 200 C 6 6 7 7 K K 303...
Page 376: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 376...
Page 377: ...1 1 0 5 in 4 CIP 2 0 75 in 5 3 Teflon 0 75 in 6 3 3 1 1 2 390 382 Mounting examples 2 377...
Page 378: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 378...
Page 382: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T 4 4 2 sc 1 2 382...
Page 383: ...3 4 15 60 14 14 15 13 ID 4 4 3 1 2 3 4 0 5 6 OK 7 OK 8 9 OK OK 10 4 4 4 383...
Page 384: ...1 2 3 1 4 5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 384...
Page 385: ...4 4 6 1 2 OK 4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus Web 5 1 2 7 25 psi 50 kPa MSDS SDS 385...
Page 387: ...3 6 3 4 4 sc sc service sc sc sc 387...
Page 389: ...6 5 6 6 300 000 300 000 100 C 0 C 379 0 4 C 0 200 C 6 6 7 389...
Page 390: ...7 Event OK K K ADC flash Flash 20 200 C 7 390...
Page 430: ...2 1 2 2 2 SC4500 2 3 1 1 1 0 5 4 2 0 75 5 3 0 75 6 430...
Page 431: ...3 3 1 1 2 442 435 2 T 2 1 T T NPT 4 7 2 T NPT 5 3 6 431...
Page 438: ...5 1 2 7 25 50 MSDS SDS 5 1 1 2 3 4 5 5 5 6 6 6 1 438...
Page 444: ......
Page 445: ......