9.
Push OK.
10.
Proceed to the calibration with a reference solution or process sample.
4.4.4 Calibration with the process sample
The sensor can remain in the process sample, or a portion of the process sample can be removed
for calibration. The reference value must be determined with a secondary verification instrument.
Note: If the sensor is calibrated for the first time, be sure to complete the zero calibration first.
1.
Select the main menu icon, then select
Devices
. A list of all of the available devices shows.
2.
Select the sensor and select
Device menu
>
Calibration
.
3.
Select
Calibration
(or
1-point value calibration
).
Note: Use the Measurement type setting to change the parameter that is calibrated.
4.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during the calibration procedure.
Hold
The sensor output value is held at the current measured value during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the controller user manual to change the
preset value.
5.
With the sensor in the process sample, push OK.
The measured value is shown.
6.
Wait for the value to stabilize and push OK.
Note: The screen may advance to the next step automatically.
7.
Measure the conductivity (or other parameter) value with a secondary verification instrument. Use
the arrow keys to enter the measured value and push OK.
8.
Review the calibration result:
• "The calibration was successfully completed."—The sensor is calibrated and ready to measure
samples. The slope and/or offset values are shown.
• "The calibration failed." —The calibration slope or offset is outside of accepted limits. Repeat
the calibration with fresh reference solutions. Clean the sensor if necessary.
9.
Push OK to continue.
10.
Return the sensor to the process and push OK.
The output signal returns to the active state and the measured sample value is shown on the
measurement screen.
4.4.5 Temperature calibration
The instrument is calibrated at the factory for accurate temperature measurement. The temperature
can be calibrated to increase accuracy.
1.
Put the sensor in a container of water.
2.
Measure the temperature of the water with an accurate thermometer or independent instrument.
3.
Select the main menu icon, then select
Devices
. A list of all of the available devices shows.
4.
Select the sensor and select
Device menu
>
Calibration
.
5.
Select
1-point temperature calibration
(or
Temperature adjustment
).
6.
Enter the exact temperature value and push OK.
7.
Return the sensor to the process.
12
English
Summary of Contents for D3433B8.99
Page 218: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 218...
Page 219: ...1 1 0 5 4 CIP 2 0 75 5 3 Teflon 0 75 6 3 3 1 1 2 232 224 2 T 219...
Page 220: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 220...
Page 224: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 14 15 13 ID 224...
Page 225: ...4 4 3 1 2 3 4 5 6 OK 7 OK 8 9 OK 10 4 4 4 1 2 3 1 4 225...
Page 226: ...5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 4 4 6 1 2 OK 226...
Page 227: ...4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus 5 1 2 7 25 psi 50 kPa MSDS SDS 5 1 Borax 227...
Page 229: ...6 3 4 4 sc sc sc sc sc 6 4 229...
Page 231: ...6 100 C 0 C 221 0 4 C 0 200 C 6 6 7 7 K K 231...
Page 290: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 290...
Page 291: ...1 1 0 5 4 CIP 2 0 75 5 3 0 75 6 3 3 1 1 2 304 296 2 291...
Page 292: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 4 1 292...
Page 296: ...4 4 50 MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 15 14 13 296...
Page 297: ...4 4 3 1 2 3 4 0 5 6 7 8 9 10 4 4 4 1 2 3 1 4 297...
Page 298: ...5 6 7 8 9 10 4 4 5 1 2 3 4 5 1 6 7 4 4 6 1 2 298...
Page 299: ...4 4 7 1 2 3 4 4 5 Modbus Modbus 5 1 2 50 MSDS SDS 5 1 299...
Page 300: ...1 2 2 3 3 4 5 5 5 2 3 6 6 6 1 6 2 299 1 2 3 3 1089 1106 23 27 C 5 5 300...
Page 301: ...6 3 4 4 sc sc sc sc sc 6 4 301...
Page 303: ...6 100 C 0 C 293 0 4 C 0 200 C 6 6 7 7 K K 303...
Page 376: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 376...
Page 377: ...1 1 0 5 in 4 CIP 2 0 75 in 5 3 Teflon 0 75 in 6 3 3 1 1 2 390 382 Mounting examples 2 377...
Page 378: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 378...
Page 382: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T 4 4 2 sc 1 2 382...
Page 383: ...3 4 15 60 14 14 15 13 ID 4 4 3 1 2 3 4 0 5 6 OK 7 OK 8 9 OK OK 10 4 4 4 383...
Page 384: ...1 2 3 1 4 5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 384...
Page 385: ...4 4 6 1 2 OK 4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus Web 5 1 2 7 25 psi 50 kPa MSDS SDS 385...
Page 387: ...3 6 3 4 4 sc sc service sc sc sc 387...
Page 389: ...6 5 6 6 300 000 300 000 100 C 0 C 379 0 4 C 0 200 C 6 6 7 389...
Page 390: ...7 Event OK K K ADC flash Flash 20 200 C 7 390...
Page 430: ...2 1 2 2 2 SC4500 2 3 1 1 1 0 5 4 2 0 75 5 3 0 75 6 430...
Page 431: ...3 3 1 1 2 442 435 2 T 2 1 T T NPT 4 7 2 T NPT 5 3 6 431...
Page 438: ...5 1 2 7 25 50 MSDS SDS 5 1 1 2 3 4 5 5 5 6 6 6 1 438...
Page 444: ......
Page 445: ......