Table 1 Sensors connected to conductivity module (continued)
Option
Description
Filter
Sets a time constant to increase signal stability. The time constant calculates the average
value during a specified time—0 (no effect, default) to 200 seconds (average of signal
value for 200 seconds). The filter increases the time for the sensor signal to respond to
actual changes in the process.
Data logger interval
Sets the time interval for sensor and temperature measurement storage in the data log—
5, 30 seconds or 1, 2, 5, 10, 15 (default), 30, 60 minutes
Reset to default
values
Sets the Settings menu to the factory default settings and resets the counters. All sensor
information is lost.
Table 2 Sensors connected to sc digital gateway
Option
Description
Name
Changes the name that corresponds to the sensor on the top of the measure screen. The
name is limited to 12 characters in any combination of letters, numbers, spaces or
punctuation.
Measurement type
Changes the measured parameter to Conductivity (default), Resistivity, TDS (total
dissolved solids) or Salinity. When the parameter is changed, all other configured
settings are reset to the default values.
Unit
Changes the units for the selected measurement—conductivity: μS/cm (default) or
mS/cm; resistivity: MΩ (default) or kΩ; TDS: ppm (default) or ppb; salinity:
0
/
00
(ppt)
Temperature
Sets the temperature units to °C (default) or °F.
Format
Changes the number of decimal places that are shown on the measurement screen to
X.XXX, XX.XX, XXX.X or XXXX (default).
Filter
Sets a time constant to increase signal stability. The time constant calculates the average
value during a specified time—0 (no effect, default) to 60 seconds (average of signal
value for 60 seconds). The filter increases the time for the sensor signal to respond to
actual changes in the process.
Data logger interval
Sets the time interval for sensor and temperature measurement storage in the data log—
Disabled, 5, 10, 15, 30 seconds, 1, 5, 10, 15 (default), 30 minutes or 1, 2, 6, 12 hours
TDS (total dissolved
solids)
Note: The TDS (total dissolved solids) setting is only available when the Measurement type setting is
set to TDS.
Sets the factor that is used to convert conductivity to TDS—NaCl (default) or User
defined entries (enter a factor between 0.01 and 99.99 ppm/µS, default: 0.492 ppm/µS).
Cell constant
parameters
Select the cell constant K.—Sets the cell constant range to 0.01, 0.05, 0.10, 0.5 (default),
1.0 HACH, 1.0 Polymetron, 5.00 or 10.0.
Cell constant—Enters the actual certified K value from the label on the sensor cable.
When the certified K value is entered, the calibration curve is defined.
T-compensation
Adds a temperature-dependent correction to the measured value—Linear (default:
2.0%/°C, 25 °C), Ammonia, Natural water, User compensation table or None.
When User compensation table is selected, the user can enter x,y (%/°C, °C) points in
ascending order.
Note: The Natural water option is not available when the Measurement type setting is set to TDS.
Temperature sensor
Sets the temperature element for automatic temperature compensation to Manual,
PT100 or PT1000 (default). If no element is used, set to Manual and set a value for
temperature compensation (default: 25 °C).
When Temperature sensor is set to PT100 or PT1000, refer to
on page 10 to set the Factor selection setting.
Note: If Temperature sensor is set to Manual and the sensor is replaced or the sensor days are reset,
the Temperature sensor setting automatically changes back to the default setting (PT1000).
English
9
Summary of Contents for D3433B8.99
Page 218: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 218...
Page 219: ...1 1 0 5 4 CIP 2 0 75 5 3 Teflon 0 75 6 3 3 1 1 2 232 224 2 T 219...
Page 220: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 220...
Page 224: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 14 15 13 ID 224...
Page 225: ...4 4 3 1 2 3 4 5 6 OK 7 OK 8 9 OK 10 4 4 4 1 2 3 1 4 225...
Page 226: ...5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 4 4 6 1 2 OK 226...
Page 227: ...4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus 5 1 2 7 25 psi 50 kPa MSDS SDS 5 1 Borax 227...
Page 229: ...6 3 4 4 sc sc sc sc sc 6 4 229...
Page 231: ...6 100 C 0 C 221 0 4 C 0 200 C 6 6 7 7 K K 231...
Page 290: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 290...
Page 291: ...1 1 0 5 4 CIP 2 0 75 5 3 0 75 6 3 3 1 1 2 304 296 2 291...
Page 292: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 4 1 292...
Page 296: ...4 4 50 MSDS SDS 4 4 1 K T T 4 4 2 sc 1 2 3 4 15 60 14 15 14 13 296...
Page 297: ...4 4 3 1 2 3 4 0 5 6 7 8 9 10 4 4 4 1 2 3 1 4 297...
Page 298: ...5 6 7 8 9 10 4 4 5 1 2 3 4 5 1 6 7 4 4 6 1 2 298...
Page 299: ...4 4 7 1 2 3 4 4 5 Modbus Modbus 5 1 2 50 MSDS SDS 5 1 299...
Page 300: ...1 2 2 3 3 4 5 5 5 2 3 6 6 6 1 6 2 299 1 2 3 3 1089 1106 23 27 C 5 5 300...
Page 301: ...6 3 4 4 sc sc sc sc sc 6 4 301...
Page 303: ...6 100 C 0 C 293 0 4 C 0 200 C 6 6 7 7 K K 303...
Page 376: ...2 1 1 2 1 2 2 2 SC4500 2 3 1 376...
Page 377: ...1 1 0 5 in 4 CIP 2 0 75 in 5 3 Teflon 0 75 in 6 3 3 1 1 2 390 382 Mounting examples 2 377...
Page 378: ...2 1 NPT 4 7 2 NPT 5 CIP 3 6 3 2 SC SC SC sc sc SC sc 4 378...
Page 382: ...4 4 7 25 psi 50 kPa MSDS SDS 4 4 1 K T 4 4 2 sc 1 2 382...
Page 383: ...3 4 15 60 14 14 15 13 ID 4 4 3 1 2 3 4 0 5 6 OK 7 OK 8 9 OK OK 10 4 4 4 383...
Page 384: ...1 2 3 1 4 5 OK 6 OK 7 OK 8 9 OK 10 OK 4 4 5 1 2 3 4 5 1 6 OK 7 384...
Page 385: ...4 4 6 1 2 OK 4 4 7 1 2 3 OK 4 OK 4 5 Modbus Modbus Web 5 1 2 7 25 psi 50 kPa MSDS SDS 385...
Page 387: ...3 6 3 4 4 sc sc service sc sc sc 387...
Page 389: ...6 5 6 6 300 000 300 000 100 C 0 C 379 0 4 C 0 200 C 6 6 7 389...
Page 390: ...7 Event OK K K ADC flash Flash 20 200 C 7 390...
Page 430: ...2 1 2 2 2 SC4500 2 3 1 1 1 0 5 4 2 0 75 5 3 0 75 6 430...
Page 431: ...3 3 1 1 2 442 435 2 T 2 1 T T NPT 4 7 2 T NPT 5 3 6 431...
Page 438: ...5 1 2 7 25 50 MSDS SDS 5 1 1 2 3 4 5 5 5 6 6 6 1 438...
Page 444: ......
Page 445: ......