Section 3
Page 59
8990startup.fm
Advanced Sampling
18-E.
Enter the desired high or low trigger point using the numeric keypad.
Press
ACCEPT
to continue.
18-F.
Enter the Deadband value or, if programming for Flow Rate Of Change
or Rainfall, enter a time interval when the flow or rainfall change must
take place (refer to
Alarm Relays Programming on page 91
).
18-G.
Enter the number of bottles (out of the total set) to set aside for upset
samples. These will be the last bottles in the tray.
18-H.
Select the method of distribution, Sample Per Bottle or
Bottles Per Sample.
18-I.
Enter the sample volume using the numeric keypad.
18-J.
Press
ACCEPT
to continue to Variable Intervals.
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING:
HIGH TRIGGER POINT:
CANCEL
00000 in.
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING
DEADBAND:
CANCEL
0.000
CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING:
NUMBER OF BOTTLES
CANCEL
2
CLEAR
ENTRY
ENTER 1 — 4
11:00 AM 21 - APR - 01
UPSET SAMPLES
ACCEPT
UPSET SAMPLING
SAMPLE VOLUME:
CANCEL
0000
CLEAR
ENTRY
ENTER 10 — 9999
Summary of Contents for 900 MAX
Page 1: ...Catalog Number 8990 Sigma 900 MAX Refrigerated Sampler INSTRUMENT MANUAL 07 03 4ed ...
Page 2: ......
Page 4: ...Visit http www hach com ...
Page 16: ...Visit http www hach com ...
Page 24: ...Visit http www hach com ...
Page 36: ...Visit http www hach com ...
Page 38: ...Visit http www hach com ...
Page 70: ...Visit http www hach com ...
Page 98: ...Visit http www hach com ...
Page 114: ...Visit http www hach com ...
Page 148: ...Visit http www hach com ...
Page 150: ...Visit http www hach com ...