Page 80
Modem
8990cm.fm
Section 6
6.2 Modem
6.2.1 Modem Connection
This connection is for interfacing the optional internal modem (Cat. No. 1602)
with a standard public telephone line.
6.2.2 Modem Programming
1.
From the Main Menu, select
OPTIONS>ADVANCED
OPTION>COMMUNICATIONS SETUP
.
Note:
There is no way of independently enabling power to just the internal modem
without also enabling the power to the Cellular Communications Option if it is
present. However, this poses no problem even if the sampler is physically
connected directly to a phone line without a cellular phone.
2.
Highlight Modem Setup using the
UP
and
DOWN
keys. Press
ACCEPT
.
3.
Enable modem power by pressing the
CHANGE CHOICE
key. Modem
power is turned off when not in use to conserve battery power.
Table 14 Modem CPU Connections (J8)
Pin
Signal Description
A
Tip
B
Ring
C
12 V dc
D
12 V dc Reference
11:00 AM 21 - APR - 01
COMUNICATION SETUP
SELECT
MODEM SETUP
RETURN
RS232 SETUP
11:00 AM 21 - APR - 01
MODEM SETUP
ACCEPT
CHANGE
CHOICE
MODEM POWER:
CANCEL
ENABLED
CHOICES: ENABLED, DISABLED
Summary of Contents for 900 MAX
Page 1: ...Catalog Number 8990 Sigma 900 MAX Refrigerated Sampler INSTRUMENT MANUAL 07 03 4ed ...
Page 2: ......
Page 4: ...Visit http www hach com ...
Page 16: ...Visit http www hach com ...
Page 24: ...Visit http www hach com ...
Page 36: ...Visit http www hach com ...
Page 38: ...Visit http www hach com ...
Page 70: ...Visit http www hach com ...
Page 98: ...Visit http www hach com ...
Page 114: ...Visit http www hach com ...
Page 148: ...Visit http www hach com ...
Page 150: ...Visit http www hach com ...