Installing the Distributor (Multiple Bottle Operation)
8990hrd.fm
Section 2
2.6.1 Distributor Arm Alignment
1.
Program the sampler for 24-bottle operation.
2.
Press
START PROGRAM
to set the distributor shaft to the
Bottle #1 position.
3.
Place the arm on the distributor shaft and align the rib on the inside wall of
the control housing skirt.
4.
Secure the arm to the shaft by tightening the
1
/
8
in. hex-head screw,
located on the distributor arm.
Figure 13
Distributor Tubing in Arm
Figure 14
Distributor Installation
1.
Distributor Shaft
2.
Nozzle End (
1
/
8
in. max)
1.
Refrigerator
2.
Distributor Assembly
3.
Distributor Arm
1
2
1
2
3