Page 69
8990options.fm
Optional Device Installation
Section 5
Optional Device Installation
This section describes how to setup a rain gauge to the Sampler as well as
how to connect, program, calibrate, and maintain the optional water
quality probes:
5.1 Rain Gauge
An external “tipping bucket” rain gauge (Cat. No. 2149) can be connected to
the Rain Gauge connector of the sampler (
Figure 21
). The rain gauge
provides a dry contact closure to the sampler.
As rainfall collects in the 20 cm (8 in.) diameter funnel, it is directed into one
side of a “tipping bucket” assembly. As each bucket fills, it causes the bucket
assembly to tip and empty out into the bottom of the rain gauge. Each tip of
the bucket causes a single contact closure to the rain gauge and sends a
short 12 V dc pulse into pin C of the Rain Gauge connector. Each pulse (tip)
represents 0.025 cm (0.01 in.) of rain.
Figure 21
Rain Gauge Tipping Bucket
•
Rain Gauge (
section 5.1 on
page 69
•
Dissolved Oxygen Probe
(
section 5.4 on page 74
)
•
pH Probe
(
section 5.2 on page 70
)
•
Conductivity Probe
(
section 5.5 on page 76
)
•
ORP Probe
(
section 5.3 on page 72
)
Table 7 Rain Gauge Base Board Connections (J5)
Pin
Signal Description
A
+12 V dc source output
B
not used
C
+12 V dc pulse input
D
not used
E
not used
F
not used
Summary of Contents for 900 MAX
Page 1: ...Catalog Number 8990 Sigma 900 MAX Refrigerated Sampler INSTRUMENT MANUAL 07 03 4ed ...
Page 2: ......
Page 4: ...Visit http www hach com ...
Page 16: ...Visit http www hach com ...
Page 24: ...Visit http www hach com ...
Page 36: ...Visit http www hach com ...
Page 38: ...Visit http www hach com ...
Page 70: ...Visit http www hach com ...
Page 98: ...Visit http www hach com ...
Page 114: ...Visit http www hach com ...
Page 148: ...Visit http www hach com ...
Page 150: ...Visit http www hach com ...