Orbisphere 510/511/512 Analyzers - Measurement Menu
45 of 138
Operator Manual
ORBISPHERE
4.1.4 Sample Mode Measurements (portable instrument)
When the numeric view is selected, the normal
measurement window is displayed.
Press the function key
(on the header bar) to initiate
the sample measurement.
The text "In progress" and the measurement value are
displayed sequentially. The measurement process stops
when the stop criteria selected are fulfilled.
Criteria Configuration (portable instrument)” on page 46
The measurement process is stopped when:
• The "stop criteria" is fulfilled, typically when the gas
concentration reaches the set threshold
• The function key
has been pressed
• The "Maximum time" criteria is reached
• An error has occurred (e.g. sensor out)
When the sample measurement stops because the stop
criteria are fulfilled, the gas concentration and
temperature are no longer refreshed. They indicate the
measurement when the stop criteria was reached. If the
channel was configured for TPO or TPA calculation, enter
the parameters (see
“TPO and TPA Parameters (portable
).
Note:
The mini graph continues to be refreshed continuously.
If the sample mode is stopped for another reason (user
abort, max. time, or measurement error), the “aborted”
message is displayed.
The reason for the stop is explained when the icon
is
pressed (as with any event).
Summary of Contents for ORBISPHERE 510
Page 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Page 2: ......
Page 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Page 144: ......