76 of 138
Inputs/Outputs Menu - Orbisphere 510/511/512 Analyzers
Operator Manual
ORBISPHERE
6.4.4 Direct Test
6.4.5 Characteristics Test
Test to check the calibration of the analog outputs. A
precision ammeter connected at the corresponding
analog output connection point is required.
Select a value (4, 12, 20 mA available) and compare
this value (+ 0.02 mA) with that which the ammeter
shows.
A calibration is required if the value on the ammeter
differs from the current selected (+ 0.02 mA).
Note:
It is possible to test one analog output without interfering with the others. During the
test, the other analog outputs will continue to indicate the measurement.
This is a test for the correct operation of the peripherals
connected to the analog outputs, by verifying that the PLC
computes the correct value.
The analog output will send the current corresponding to
the value entered in the text boxes.
Type in a test value for the analog output to be tested,
and check for the related action on the peripheral.
Summary of Contents for ORBISPHERE 510
Page 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Page 2: ......
Page 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Page 144: ......