84 of 138
Communication Menu - Orbisphere 510/511/512 Analyzers
Operator Manual
ORBISPHERE
EC Gas Sensor Calibration Report Example
Calibration report nb 1
Type. . . . . . . . . . . . . . . . . . . . . Sensor calibrated on this instrument
Date (yy.mm.dd - hh:mm) . . . . 05.02.17 - 18:40
Operator . . . . . . . . . . . . . . . . . jp
Operator ID . . . . . . . . . . . . . . . 3
Calibration coefficient . . . . . . . 122.40 µA/bar
Ratio with ideal membrane . . . 92.72 %
Ratio with last calibration. . . . . 97.35 %
Stability . . . . . . . . . . . . . . . . . . 0.000 %
Calibration mode . . . . . . . . . . . Direct value
Calibration value . . . . . . . . . . . 0.750000
Gas unit . . . . . . . . . . . . . . . . . . bar
Gas phase . . . . . . . . . . . . . . . . Liquid
Liquid . . . . . . . . . . . . . . . . . . . . Water
Interferences . . . . . . . . . . . . . . Disable
Temperature . . . . . . . . . . . . . . 20.1 °C
Barometric pressure . . . . . . . . 1.020 bar
External pressure . . . . . . . . . . 0.976 bar
Sensor current . . . . . . . . . . . . . 79.857 µA
Smart EC Gas Sensor Connection Calibration Report Example
Calibration report nb 1
Type. . . . . . . . . . . . . . . . . . . . . Sensor calibrated on another instrument
Date (yy.mm.dd - hh:mm) . . . . 05.01.17 - 15:30
Operator . . . . . . . . . . . . . . . . . Armstrong
Calibration coefficient . . . . . . . 62.75 µA/bar
Connection date . . . . . . . . . . . 05.03.15 - 12:26
Serial number . . . . . . . . . . . . . 1
Membrane . . . . . . . . . . . . . . . . 2956A
TC Gas Sensor Calibration Report Example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . 06.02.18 - 15:20
Operator . . . . . . . . . . . . . . . . . jp
Operator ID . . . . . . . . . . . . . . . 3
Calibration coefficient . . . . . . . 18.23 mV/bar
Ratio with ideal membrane . . . 107.22 %
Ratio with last calibration. . . . . 106.32 %
Stability . . . . . . . . . . . . . . . . . . 0.337 %
Calibration value . . . . . . . . . . . 965.000000
Gas unit . . . . . . . . . . . . . . . . . . mbar
Gas phase . . . . . . . . . . . . . . . . Gas
Liquid . . . . . . . . . . . . . . . . . . . . Water
Interferences . . . . . . . . . . . . . . Disable
Temperature . . . . . . . . . . . . . . 26.44 °C
Barometric pressure . . . . . . . . 0.962 bar
External pressure . . . . . . . . . . 0.976 bar
Sensor slope . . . . . . . . . . . . . . 18.797 mV/s
Summary of Contents for ORBISPHERE 510
Page 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Page 2: ......
Page 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Page 144: ......