88 of 138
Communication Menu - Orbisphere 510/511/512 Analyzers
Operator Manual
ORBISPHERE
7.2 PROFIBUS-DP Communication (optional)
7.2.1 Installation
On the Orbisphere CD, there is an “Orbi2079.gsd” and an “Orbi2079.bmp” file available
in the “Profibus DP” folder to help configure the PROFIBUS-DP. The GSD file contains
the following elements:
• A module to decode the barometric pressure value and unit
• A module to convert the channel measurement data such as gas concentration,
gas unit, temperature, temperature unit and the events.
WARNING
Installation should be performed exclusively by personnel specialized and authorized to
work on electrical installations, in accordance with relevant local regulations. Disconnect
the power supply of the instrument before carrying out any work inside the instrument.
CAUTION:
Proper ESD (electrostatic discharge) protocols must be followed to prevent damage to
the product. All fittings must be properly seated and tightened to prevent any water and
dust ingress.
1) Install the PROFIBUS-DP module and the jumper J3 on the main board (location
highlighted in the illustration below).
Summary of Contents for ORBISPHERE 510
Page 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Page 2: ......
Page 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Page 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Page 144: ......