REMOTE
CONTROL
103
rout:scan?
Returns currently scanned channels.
Parameter: start channel followed by stop channel
Secondary display: MEASure2 command
meas2:volt:dc?
Returns DC Voltage value on the 2
nd
display.
meas2:volt:ac?
Returns AC Voltage value on the 2
nd
display.
meas2:curr:dc?
Returns DC Current value on the 2
nd
display.
meas2:curr:ac?
Returns AC Current value on the 2
nd
display.
meas2:res?
Returns 2W Resistance value on the 2
nd
display.
meas2:fres?
Returns 4W Resistance value on the 2
nd
display.
meas2:freq?
Returns Frequency value on the 2
nd
display.
meas2:per?
Returns Period value on the 2
nd
display.
meas2:temp?
Returns Temperature value on the 2
nd
display.
Secondary display: CONFigure2 command
conf2:volt:dc Configure
2
nd
display to DC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:dc_1 (DC Voltage, 1V range)
conf2:volt:ac Configure
2
nd
display to AC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:ac_1 (AC Voltage, 1V range)
conf2:curr:dc Configure
2
nd
display to DC Current.
Parameter: NR2, min, max
Example: conf2:curr:dc_10e-3 (DC Current, 10mA range)
conf2:curr:ac Configure
2
nd
display to AC Current.
Parameter: NR2, min, max
Example: conf2:curr:ac_10e-3 (AC Current, 10mA range)
conf2:res Configure
2
nd
display to 2W Resistance.
Parameter: NR2, min, max
Example: conf2:res_10e2 (2W Resistance, 1k
Ω
range)
conf2:fres Configure
2
nd
display to 4W Resistance.
Parameter: NR2, min, max
Example: conf2:fres_10e2 (Resistance, 1k
Ω
range)
conf2:freq Configure
2
nd
display to Frequency.
conf2:per Configure
2
nd
display to Period.
conf2:temp Configure
2
nd
display to Temperature.
GDM-8200 Series User Manual
104
F
AQ
●
What is the Output key used for?
●
What are the V and I mark above the Up/Down key used for?
●
I pressed the EXIT key but cannot get out of Scanner mode.
●
The GDM-8200 series performance does not match the specification.
What is the Output key used for?
The Output key is reserved for future use, in the optional
Power Source module.
What are the V and I mark above the Up/Down key used for?
The V/I mark is reserved for future use, in the optional
Power Source module.
I pressed the EXIT key but cannot get out of Scanner mode.
Press the EXIT key, followed by the ACV (Scan) or
DCV (Step) key.
The GDM-8200 series performance does not match the specification.
Make sure the device is powered On for at least 30
minutes, 18°C~+28°C. This is necessary to
stabilize the unit to match the specification.
If there is still a problem, please contact your local dealer or GWInstek at
.