ADVANCED
MEASUREMENT
45
Max/Min Measurement
Applicable to
Background
Maximum and Minimum measurement stores the highest
(maximum) or lowest (minimum) reading and shows it on
the 2nd display.
For Max measurement, press the
MX/MN key once.
1. Activate
Max/Min
For Min measurement, press the
MX/MN key twice.
2. Max (Min)
result appears
MIN (MAX) Indicates Min (Max) measurement
0.11516
2nd display shows the Min (Max)
measurement result
Deactivate
Max/Min
measurement
To cancel the Max/Min
measurement, press the MX/MN
key for 2 seconds, or simply activate
another measurement.
GDM-8200 Series User Manual
46
Relative Value Measurement
Applicable to
Background
Relative measurement stores a value, typically the data at
the moment, as the reference. The following
measurement is shown as the delta between the reference.
1. Activate
Relative
measurement
Press the REL key. The
measurement reading at the moment
becomes the reference value.
2. Relative
measurement
display appears
REL
Indicates Relative value measurement
2nd
display
Shows the reference value
1st
display
Shows the delta between the current
measurement data and the reference value
Manually set the
reference value
1.
To set the reference value
manually, press the Shift key
followed by the REL key. The
setting appears.
REL
Indicates Relative measurement
1st
display
Shows the reference value
2nd
display
Indicates Relative value modification
2.
Use the Left/Right key to move
the flashing point (cursor), and
use the Up/Down key to change
the value.