SCANNER
(OPTIONAL)
75
Scanner Configuration Record
Channel Wire
color
Measure type
Note
CH1 H
L
CH2 H
L
CH3 H
L
CH4 H
L
CH5 H
L
CH6 H
L
CH7 H
L
CH8 H
L
CH9 H
L
CH10 H L
CH11 H L
CH12 H L
CH13 H L
CH14 H L
CH15 H L
CH16 H L
CH17 H L
CH18 H L
CARD INPUT H
L
CARD SENSE H
L
AMPS H L
GDM-8200 Series User Manual
76
Setup Scan
Overview
Scan type
Simple
Sets the scanned channel range, loop
count, and timer length. All channels
have a common measurement item.
Advanced
In addition to the above Simple Scan
setting, allows custom setting for each
channel, such as measurement item,
range, and rate.
Timer setting
Sets the duration between each scan loop (Scan
operation) or between each scanned channel (Step
operation).
Count setting
Sets the number of scan operation (loop).
Trigger setting
Internal
(Continuous)
GDM-8200 series keeps triggering
continuously until the scan reach the end
of loop count. Then it goes into the idle
mode.
External
(Manual)
GDM-8200 series stays in the idle mode
by default. The trigger timing is manually
controlled by the user from the front
panel (TRIG key).
Scan operation
Scan
Measures all specified channel range at
each trigger event. Timer setting (page77)
applies between each scan (the whole
channel range).
Step
Measures a single channel in the specified
range at each trigger event. Timer setting
(page77) applies between each channel.
Monitor
Selects just one channel and continuously
measure it.