SCANNER
(OPTIONAL)
83
Recall Scan/Step result
Panel operation
1.
After the Scan/Step is
completed, the data is stored
internally. Press the Shift key
followed by the ACI (Recall) key.
2.
The first channel appears. (example: channel 101)
3.
To view the Max/Min/Average
data, press the Left key.
4.
To move to the next channel,
press the Up/Down key.
5.
Press the EXIT key to get out
from recall mode.
Setup and run monitoring
Panel operation
1.
Press the Shift key, the 2
ND
key
(MENU), the Left key twice.
The Scan menu appears.
2.
Press the Down key followed by
the Left key. The Monitor Scan
setting menu appears.
GDM-8200 Series User Manual
84
3.
Press the Down key. The
channel selection appears.
4.
Move the cursor to the channel
using the Left/Right key, and
change the value using the
Up/Down key.
5.
When finished, press the
ENTER key. The Count setting
appears.
Range
1 ~ 128
* If the entered value is out of
the range, an error message
appears (Small or Large).
6.
When finished, press the
ENTER key. The Monitoring
starts.