Section 4
OM6530-C1-00
19 November, 2019
4-37
Pressing the
<Thresholds>
key will display the coefficients stored for each
threshold. By pressing the
<Edit>
key the numbers may be altered by pressing a
new sequence of numbers followed by pressing the
<Ok>
key. Pressing the
<PREVIOUS>
key exits to the SOFCAL Calibrate Menu.
4.9.2.2.3.Test Volts
Pressing the
<Test Volts>
key will display the test voltages and the coefficients
stored for each step. A new value may be entered for each test voltage by
pressing the <Edit> key and the numeric keys for the new value, followed by the
<Ok>
key. Press the UP or DN key to select the next test voltage coefficient. The
different test voltages are: (1; 3; 10; 30; 100; 300; 1000) V in both polarities (±).
To determine the output test voltage coefficient a DC Voltmeter must connected
to the Output/Source connector. The following equation is used to calculate the
value to enter for each output test voltage range. (see section 6.1.2)
Test number to DC V reading - | Nominal Voltage |
Voltage = be keyed =
___________________________________________
x 10